We report our experience with 215 recalled neck modular stems due to corrosion. Among the 195 hips with 2 years follow-up, 56% had no clinical symptoms, 26% had groin pain (typical of corrosion), and 17% had other symptoms. Cobalt levels were comparable between asymptomatic (3.4 mu g/L, range 0.7-7.3 mu g/L) and symptomatic patients (4.0 mu g/L range 0-13.2 mu g/L). Abnormal imaging findings were seen in 46% of symptomatic and 11% of asymptomatic hips (P = 0.001). Twenty-six hips (13%) have either undergone revision surgery or have been scheduled. Evidence of corrosion was seen at revision surgery in all patients. Despite modest elevations in serum cobalt levels, abnormal imaging studies were seen in 36%, clinical symptoms were seen in 44%, and revision for corrosion was undertaken or scheduled in 13% of the hips. (C) 2014 Elsevier Inc. All rights reserved.
机译:我们报告了我们因腐蚀而召回的215个召回颈部模块化阀杆的经验。在进行了2年随访的195例髋关节中,没有临床症状的占56%,有腹股沟痛(典型的是腐蚀)的占26%,有其他症状的占17%。无症状(3.4μg/ L,范围0.7-7.3μg/ L)和有症状患者(4.0μg/ L范围0-13.2μg/ L)之间的钴水平相当。在有症状的46%和无症状的髋部中发现异常的影像学发现(P = 0.001)。 26髋(13%)接受了翻修手术或已安排好手术时间。在所有患者的翻修手术中均发现了腐蚀迹象。尽管血清钴水平适度升高,但在36%的患者中发现了异常的影像学研究,在44%的患者中发现了临床症状,在13%的髋关节中进行或计划进行腐蚀修复。 (C)2014 Elsevier Inc.保留所有权利。