首页> 外文期刊>The Journal of arthroplasty >Early failure of articular surface replacement XL total hip arthroplasty.
【24h】

Early failure of articular surface replacement XL total hip arthroplasty.

机译:关节表面置换的早期失败XL全髋关节置换术。

获取原文
获取原文并翻译 | 示例
           

摘要

The ASR (articular surface replacement) XL (DePuy, Warsaw, Ind) metal-on-metal hip arthroplasty offers the advantage of stability and increased motion. However, an alarming number of early failures prompted the evaluation of patients treated with this system. A prospective study of patients who underwent arthroplasty with the ASR XL system was performed. Patients with 2-year follow-up or any revision were included. Failure rates, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores, and radiographs were evaluated. Ninety-five patients (105 hips) were included. There were 16 revisions. Thirteen (12%) were aseptic acetabular failures. Eight were revised for aseptic loosening; 4, for metallosis; 1, for malposition; 2, for infection; and 1, for periprosthetic fracture. Mean time to revision was 1.6 years (0.18-3.4 years). The ASR XL with a revision rate of 12% is the second reported 1 piece metal-on-metal system with a significant failure rate at early follow-up. This particular class of implants has inherent design flaws that lead to early failure.
机译:ASR(关节表面置换)XL(DePuy,华沙,印度)金属对金属髋关节置换术具有稳定性和增加运动的优势。但是,数量惊人的早期失败促使对使用该系统治疗的患者进行评估。对使用ASR XL系统进行关节置换术的患者进行了一项前瞻性研究。包括2年随访或任何修订的患者。评估失败率,西安大略省和麦克马斯特大学骨关节炎指数(WOMAC)得分以及X光片。包括九十五例患者(105髋)。有16个修订版本。十三例(12%)是无菌性髋臼衰竭。修订了八项无菌松动; 4,用于金属化; 1,错位; 2,用于感染; 1,假体周围骨折。平均修订时间为1.6年(0.18-3.4年)。修订率为12%的ASR XL是第二个报告的1件式金属对金属系统,在早期随访中具有显着的故障率。这类特殊的植入物具有固有的设计缺陷,会导致早期失败。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号