Factory-assembled polyethylene linings in metal backing show higher wear rates compared to equisized all-polyethylene sockets.

机译:与同等的全聚乙烯插座相比,在金属背衬中工厂组装的聚乙烯衬里具有更高的磨损率。

获取原文
获取原文并翻译 | 示例
           

摘要

Two series of cementless total hip arthroplasty with acetabular sockets of a threaded truncated cone design were compared regarding volumetric wear rates. The first series included all-polyethylene acetabular sockets of the type Endler (E-PE); in the second series, a nonmodular titanium metal-backed polyethylene (E-MB) socket with an identical outer shape to E-PE was implanted. Bearings were articulated with alumina 32-mm diameter ball heads. Ninety-five retrieved devices were examined with a modified fluid displacement method using dental self-polymerizing precision casts. The sockets had to be revised mainly because of wear-induced osteolysis: E-PE after 10.6 years on average and E-MB after 7.8 years (P = .002). Comparison with unused sockets showed mean wear rates of 63 mm(3)/y for E-PE and 120 mm(3)/y for E-MB (P = .0008). Increased contact stress and load deformation due to reduction of polyethylene thickness in E-MB compared to E-PE were identified as predominant reasons for higher wear rates.
机译:比较了两个系列的非截骨全髋关节置换术和带螺纹截头圆锥设计的髋臼窝的容积磨损率。第一个系列包括Endler(E-PE)型的全聚乙烯髋臼承窝;在第二系列中,植入了具有与E-PE相同外形的非模块化钛金属背衬聚乙烯(E-MB)插座。轴承用直径32毫米的氧化铝球头铰接。使用改良的流体置换方法,使用牙科自聚合精密铸模检查了九十五个回收装置。必须对承窝进行修订,主要是因为磨损引起的骨溶解:平均E-PE在10.6年后使用,E-MB在7.8年后使用(P = .002)。与未使用的插座进行比较显示,E-PE的平均磨损率为63 mm(3)/ y,E-MB的平均磨损率为120 mm(3)/ y(P = .0008)。与E-PE相比,由于E-MB中聚乙烯厚度的减少,增加了接触应力和载荷变形,这被认为是造成较高磨损率的主要原因。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号