We report and analyze the causes of a fracture in a proximally modular, distally tapered fluted MP stem in a 48-year-old woman (168 cm, 67 kg) with severe proximal bone deficiency. Evidence of fatigue failure with striations initiated laterally was observed in the laser etching of the tensile aspect of the prosthesis. However, metallurgical analysis suggested that laser engraving did not alter the microstructure of the stem. Stress due to the absence of proximal femoral bone support may have been sufficiently high to put this particular stem at risk for fatigue fracture. This important complication should be addressed when choosing this therapeutic option in cases with substantial proximal femoral bone loss. Strut allograft support should be recommended in such cases.
机译:我们报告并分析了一个严重的近端骨缺损的48岁妇女(168 cm,67 kg)的近端模块化,远端渐缩的带凹槽的MP茎中骨折的原因。在假体拉伸方向的激光蚀刻中,观察到了疲劳疲劳和横向条纹现象。但是,冶金分析表明,激光雕刻不会改变杆的微观结构。由于缺乏股骨近端支撑而产生的压力可能已经足够高,以致于使该特定茎具有疲劳断裂的风险。在股骨近端大量骨丢失的情况下选择这种治疗选择时,应解决这一重要并发症。在这种情况下,应推荐支撑同种异体支撑。