首页> 外文期刊>The Journal of arthroplasty >Minimum 5-year results in 96 consecutive hips treated with a tapered titanium stem system.
【24h】

Minimum 5-year results in 96 consecutive hips treated with a tapered titanium stem system.

机译:使用锥形钛杆系统治疗的连续96个髋关节至少获得5年结果。

获取原文
获取原文并翻译 | 示例
           

摘要

We performed total hip arthroplasty with single titanium stem in 96 consecutive, nonselected hips. All patients, regardless of bone type and shape, neck shaft angulation, or age, received the same implant. Patients were followed for a minimum of 5 years, and an independent radiographer evaluated the hips for ingrowth, subsidence, leg-length discrepancy, and remodeling. The average Harris Hip score was 96 points (range, 73-100 points) at final follow-up. Radiographically, all stems were ingrown. No stem had more than 3 mm of subsidence, and there were no leg-length discrepancies more than 5 mm. We concluded that this titanium stem is a versatile option for total hip arthroplasty in patients with a wide variety of demographic and femur characteristics.
机译:我们在96个连续的,未选择的髋关节中用单个钛杆进行了全髋关节置换术。所有患者,无论骨骼类型和形状,颈干轴的角度或年龄如何,都接受相同的植入物。对患者进行了至少5年的随访,一名独立的放射线照相师评估了髋关节的向内生长,下陷,腿长差异和重塑。在最后的随访中,Harris Hip的平均得分为96分(范围73-100分)。影像学上,所有茎都向内生长。没有茎具有超过3毫米的下沉,并且腿长差异没有超过5毫米。我们得出的结论是,对于具有多种人口统计学特征和股骨特征的患者,此钛杆是全髋关节置换术的多功能选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号