首页> 外文期刊>The international journal of artificial organs >Aortic root compliance influences hemolysis in mechanical heart valve prostheses: An in-vitro study
【24h】

Aortic root compliance influences hemolysis in mechanical heart valve prostheses: An in-vitro study

机译:体外研究表明,主动脉根顺应性影响机械性心脏瓣膜假体的溶血

获取原文
获取原文并翻译 | 示例
           

摘要

Mechanical heart valve prostheses are known to activate coagulation and cause hemolysis. Both are particularly dependent on the leaflet dynamics, which in turn depends on the flow field in the aortic root influenced by the aortic root geometry and its compliance. Compliance reduction of large vessels occurs in aging patients, both in those who have atherosclerotic diseases and those who do not. In this study we investigated the correlation between hemolysis and the compliance of the proximal aorta in a novel, pulsatile in vitro blood tester using porcine blood. Two mechanical heart valves, the St Jude Medical (SJM) bileaflet valve and a trileaflet valve prototype (Triflo) were tested for hemolysis under physiological conditions (120/80 mm Hg, 4.5 l/min, 70 bpm) and using two different tester setups: with a stiff aorta and with a compliant aorta. Valve dynamics were subsequently analyzed via high-speed videos. In the tests with the Triflo valve, the free plasma hemoglobin increased by 13.4 mg/dl for the flexible and by 19.3 mg/dl for the stiff setup during the 3-hour test. The FFT spectra and closing speed showed slight differences for both setups. Free plasma hemoglobin for the SJM valve was up by 22.2 mg/dl in the flexible and 42.7 mg/dl in the stiff setup. Cavitation induced by the higher closing speed might be responsible for this, which is also indicated by the sound spectrum elevation above 16 kHz.
机译:已知机械心脏瓣膜假体可激活凝血并引起溶血。两者都特别取决于小叶动力学,而小叶动力学又取决于受主动脉根几何形状及其顺应性影响的主动脉根中的流场。在患有动脉粥样硬化疾病的人和没有动脉粥样硬化疾病的人的衰老患者中,大血管的顺应性降低。在这项研究中,我们研究了使用猪血的新型脉冲式体外血液测试仪中溶血与近端主动脉顺应性之间的相关性。在生理条件下(120/80 mm Hg,4.5 l / min,70 bpm)并使用两种不同的测试仪设置,测试了两个机械心脏瓣膜,St Jude Medical(SJM)双叶瓣和三叶瓣原型(Triflo)的溶血情况:具有坚硬的主动脉和顺应性主动脉。随后通过高速视频分析阀门动态。在使用Triflo阀的测试中,在3小时的测试中,柔性血浆的游离血浆血红蛋白增加了13.4 mg / dl,而硬质设置的血浆游离血红蛋白增加了19.3 mg / dl。两种设置的FFT频谱和关闭速度均显示出细微差别。 SJM阀的游离血浆血红蛋白在柔性装置中增加了22.2 mg / dl,在刚性装置中增加了42.7 mg / dl。较高的关闭速度引起的气蚀可能是造成这种情况的原因,这也可以由高于16 kHz的声谱高度表示。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号