Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.

机译:植入式心脏复律除颤器的声音警报可检测到导体间断。

获取原文
获取原文并翻译 | 示例
           

摘要

The lead conductor integrity of implantable cardioverter defibrillator devices is inferred from impedance measurements; however, intermittent discontinuity can be difficult to detect or confirm. Newer devices can perform daily lead impedance self-testing, and some even have audible alarms that promptly warn patients of anomalies. In the present case, the audible alarms were solely responsible for the timely identification of an intermittent, otherwise clinically nonreproducible, form of potentially fatal implantable cardioverter defibrillator system failure.
机译:植入式心脏复律除颤器设备的导线完整性是通过阻抗测量得出的。但是,间歇性的间断可能很难检测或确认。较新的设备可以执行日常的导线阻抗自检,有些甚至具有声音警报,可以及时警告患者异常情况。在当前情况下,声音警报仅负责及时识别可能致命的可植入式心脏复律除颤器系统故障的间歇性(否则在临床上不可重现)形式。