首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >ROUTINE LEAD ISOTOPE DETERMINATIONS USING A LEAD-207 LEAD-204 DOUBLE SPIKE - A LONG-TERM ASSESSMENT OF ANALYTICAL PRECISION AND ACCURACY
【24h】

ROUTINE LEAD ISOTOPE DETERMINATIONS USING A LEAD-207 LEAD-204 DOUBLE SPIKE - A LONG-TERM ASSESSMENT OF ANALYTICAL PRECISION AND ACCURACY

机译:使用LEAD-207 LEAD-204双峰的常规铅同位素测定-分析精度和准确性的长期评估

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Lead-isotope data obtained on a multicollector mass spectrometer over a four year period using a Pb-207-Pb-204 double spike to correct for the effects of mass discrimination, are reported. Considerable improvements in both precision end accuracy over conventional correction procedures were noted, without recourse to rigorous loading or run conditions. An external precision in Pb-206/Pb-204, Pb-207/Pb-204 and Pb-208/Pb-204 ratios +/-0.003, 0.003 and 0.01 (2 x standard deviation), respectively, is routinely obtainable independent of minor variations in loading and run parameters. [References: 21]
机译:报告了在多收集器质谱仪上使用Pb-207-Pb-204双尖峰在四年时间内获得的铅同位素数据,以校正质量歧视的影响。在不依靠严格的负载或运行条件的情况下,与传统的校正程序相比,在两个精度端部精度上都取得了显着的改进。通常可以独立于Pb-206 / Pb-204,Pb-207 / Pb-204和Pb-208 / Pb-204之比+/- 0.003、0.003和0.01(2 x标准偏差)获得外部精度加载和运行参数的细微变化。 [参考:21]