首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >Determination of the minimum allowable operating temperature of stationary phases in capillary columns by inverse gas chromatography
【24h】

Determination of the minimum allowable operating temperature of stationary phases in capillary columns by inverse gas chromatography

机译:反相气相色谱法测定毛细管色谱柱中固定相的最低允许工作温度

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A method was developed to determine the minimum allowable operating temperature (MiAOT) of wall coated open tubular capillary columns. Two polyethylene glycol and fourteen polysiloxanes phases with different side groups ( methyl, phenyl, cyanopropyl, trifluoropropyl, n-octyl) and backbone stiffening units (tetramethyl-p-silphenylene, tetramethyl-p, p '-sildiphenylene ether, carborane) were investigated by inverse gas chromatography. A sigmoidal profile of temperature versus column efficiency was found for almost all phases, as the column efficiency increased with temperature. The MiAOT was defined as that temperature where the column efficiency is half of its original value at elevated temperatures. It was found that the MiAOT of a stationary phase is approx. 60 K higher than its glass transition temperature.
机译:开发了一种方法来确定壁涂式开放式管状毛细管柱的最低允许工作温度(MiAOT)。研究了两个带有不同侧基的聚乙二醇和十四个聚硅氧烷相(甲基,苯基,氰基丙基,三氟丙基,正辛基)和骨架刚性单元(四甲基-对-亚苯基,四甲基-对,p'-硅二亚苯基醚,碳硼烷)反相气相色谱。随着柱效随温度的升高,几乎所有相的温度与柱效的关系呈S形。 MiAOT定义为在较高温度下柱效为其原始值的一半的温度。已经发现,固定相的MiAOT大约为1。比其玻璃化转变温度高60K。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号