【24h】

Performance of a multiplexed chemical ionization miniature cylindrical ion trap array mass spectrometer

机译:多重化学电离微型圆柱离子阱阵列质谱仪的性能

获取原文
获取原文并翻译 | 示例
           

摘要

The performance of a multiplexed array of miniature cylindrical ion trap (CIT) mass spectrometers composed of four identical sets of ion source/mass analyzer/detector channels is evaluated. Each miniature cylindrical ion trap (inner radius 2.5 mm) was coupled to an external chemical ionization/electron ionization (CI/EI) source and a miniature (inner diameter 2 mm) electron multiplier. The sample introduction system permits delivery of samples and reagent gases directly into any one or more of the ionization volumes. Experimental conditions were optimized as required for CI to occur in one or more ion volumes. The ion source potential was observed to affect the energy of [M + H](+) ions generated by self-chemical ionization in the external ion sources and, thus, the degree of fragmentation observed in resulting mass spectra. The versatility of this novel instrumentation permits two types of multi-channel experiments. In the first, different samples are analyzed at the same time in identical, parallel channels for high-throughput mass analysis using CI. In the second, the instrument is used for improved specificity in the analysis of a single sample by simultaneous measurements made with multiple methods of ionization, including EI and methane CI in adjacent channels of the instrument. [References: 38]
机译:评估了由四组相同的离子源/质量分析器/检测器通道组成的微型圆柱离子阱(CIT)质谱仪的多阵列阵列的性能。每个微型圆柱形离子阱(内径2.5毫米)都与外部化学电离/电子电离(CI / EI)源和微型(内径2毫米)电子倍增器耦合。样品引入系统允许将样品和反应气直接输送到任何一个或多个电离容积中。根据需要优化实验条件,以使CI在一个或多个离子体积中发生。观察到离子源电势会影响外部离子源中通过自化学电离产生的[M + H](+)离子的能量,因此会影响所得质谱图中的碎裂程度。这种新型仪器的多功能性允许两种类型的多通道实验。首先,使用CI在同一平行通道中同时分析不同样品,以进行高通量质量分析。在第二种方法中,该仪器可通过同时测量多种电离方法(包括仪器相邻通道中的EI和甲烷CI)进行的同时测量来提高单个样品分析的特异性。 [参考:38]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号