首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >Direct detection of large fat-soluble biomolecules in solution using membrane inlet mass spectrometry and desorption chemical ionization
【24h】

Direct detection of large fat-soluble biomolecules in solution using membrane inlet mass spectrometry and desorption chemical ionization

机译:使用膜入口质谱和解吸化学电离直接检测溶液中的大型脂溶性生物分子

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This paper presents the first membrane inlet mass spectrometry system capable of detecting large biomolecules, such as testosterone (M-r 288), testosterone acetate (M-r 330) and alpha-tocopherol (M-r 430, vitamin E). The result was obtained using a home-made chemical ionization ion source with a thermostated tubular silicone membrane mounted right in the centre of a methane CI plasma. The liquid sample was flushed through the inside of the membrane for a period of 20-25 min, where the analyte diffused into the membrane. Following this trapping period the analyte was released from the membrane into the mass spectrometer by the combined action of heat radiation from the filament and charge transfer from the chemical ionization plasma. As a result of this stimulated desorption a good desorption peak was obtained as the analyte vaporized out of the membrane. Retinol (M-r 286, vitamin A), cholecalciferol (M-r 384, vitamin D-3) and cholesterol (M-r 386) were also detected. However, these compounds (all containing a long hydrocarbon chain and being aliphatic alcohols) did not give a protonated molecule. They gave a series of cluster ions with the dominant located 20 mass units below the molecular ion. The detection limits of the new desorption chemical ionization MIMS technique were at low or sub-micromolar concentrations (high ppb levels) and the reproducibility was within 20%, when the area of the desorption peak was used for quantitation. [References: 14]
机译:本文介绍了第一个能够检测大型生物分子的膜入口质谱系统,例如睾丸激素(M-r 288),乙酸睾丸酮(M-r 330)和α-生育酚(M-r 430,维生素E)。使用自制的化学电离离子源获得的结果,该离子源具有安装在甲烷CI等离子体中心的恒温管状硅胶膜。液体样品通过膜内部冲洗20-25分钟,分析物在膜中扩散。在这个捕获期之后,通过来自灯丝的热辐射和来自化学电离等离子体的电荷转移的联合作用,将分析物从膜释放到质谱仪中。由于该刺激的解吸,当分析物从膜中汽化时,获得了良好的解吸峰。还检测到了视黄醇(M-r 286,维生素A),胆钙化固醇(M-r 384,维生素D-3)和胆固醇(M-r 386)。但是,这些化合物(均含有长烃链且为脂肪族醇)均未得到质子化分子。他们给出了一系列簇离子,其主要分子位于分子离子以下20质量单位处。当使用解吸峰面积进行定量分析时,新的解吸化学电离MIMS技术的检测限为低或亚微摩尔浓度(高ppb水平),重现性在20%以内。 [参考:14]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号