首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >Membrane solubilization with tetramethylammonium hydroxide for the preconcentration and electrothermal atomic absorption spectrometric determination of trace amounts of arsenic in water
【24h】

Membrane solubilization with tetramethylammonium hydroxide for the preconcentration and electrothermal atomic absorption spectrometric determination of trace amounts of arsenic in water

机译:氢氧化四甲铵对膜的溶解度用于预浓缩和电热原子吸收光谱法测定水中的痕量砷

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Solubilization of a mixed cellulose ester membrane filter (MF) with tetramethylammonium hydroxide (TMAH) is proposed for the preconcentration and electrothermal atomic absorption spectrometric (ETAAS) determination of trace amounts of arsenic in water. Arsenic at not more than 0.4 mu g in 100 ml of sample solution was retained on the MF by filtration as an ion associate of arsenomolybdate and tetraphenylphosphonium ions. The ion associate was dissolved in a small volume of TMAH together with the MF. After being made up to 2 ml with water, the arsenic in the concentrate was determined by ETAAS in the presence of zirconyl nitrate as a chemical modifier. This method is very simple and rapid. The detection limit, defined as three times the standard deviation of the blank, was 0.04 mu g l(-1). Inorganic matrix components in river waters, Na+, K+, Ca2+, Mg2+, SO42-, NO3- and silicate at high concentrations did not interfere with the determination. Trace components, phosphate at 0.3 mg l(-1), dodecyl sulfate at 14 mg l(-1) and aluminium at 5 mg l(-1), also did not interfere with the determination. The proposed method was applied to the analysis of river water samples. [References: 15]
机译:提出用氢氧化四甲基铵(TMAH)溶解混合纤维素酯膜滤器(MF)进行预浓缩和电热原子吸收光谱法(ETAAS)测定水中的痕量砷。通过过滤将砷在100毫升样品溶液中的浓度不超过0.4克作为砷钼酸根和四苯基phosph离子的离子缔合体保留在MF上。离子缔合物与MF一起溶解在少量的TMAH中。用水补足至2 ml后,在硝酸氧锆作为化学改性剂的情况下,通过ETAAS测定浓缩物中的砷。这种方法非常简单快捷。检出限定义为空白标准偏差的三倍,为0.04μg l(-1)。河流水中的无机基质成分,高浓度的Na +,K +,Ca2 +,Mg2 +,SO42-,NO3-和硅酸盐不会干扰测定。微量组分,磷酸盐为0.3 mg l(-1),十二烷基硫酸盐为14 mg l(-1)和铝为5 mg l(-1),也不会影响测定。将该方法应用于河流水样分析。 [参考:15]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号