Chromium(III) hexacyanoferrate(II)-based chemical sensor for the cathodic determination of hydrogen peroxide

机译:基于六氰合铁酸铬(III)的阴极过氧化氢化学传感器

获取原文
获取原文并翻译 | 示例
           

摘要

A cathodic scheme to measure hydrogen peroxide by utilizing a membrane free chromium(III) hexacyanoferrate(II) based chemical sensor is described. The cluster is prepared simply by generating a chromium(III) hexacyanoferrate(II) cluster electrochemically on a rotating disk glassy carbon electrode. Subsequently, the hydrogen peroxide can be measured at 0 V versus Ag/AgCl. This approach significantly reduces interferences, especially those by easy oxidizable compounds such as catechol and ascorbic acid. Also, no significant oxygen interference is observed at this working potential, Several electrolytes mere carefully investigated. The calibration curve was linear up to 1.3 mM (r = 0.9992) with a typical response of the sensor of about 5.6 s with injection of 0.05 mM hydrogen peroxide. The detection limit of this chemical sensor is 3.0 X 10(-8) M (S/N = 3) with a low-pass filter (time constant 0.1 s), The sensitivity of the sensor is 11.52 X 10(-6) A l mmol(-1) mm(-2) (r = 0.9992). The RSD of this sensor is 1.1%. [References: 30]
机译:描述了一种阴极方案,该方案通过使用基于膜的六价铁氰化铬(III)的化学传感器来测量过氧化氢。通过在旋转盘状玻璃碳电极上电化学生成六氰合铁酸铬(III)簇,可以简单地制备簇。随后,可以在相对于Ag / AgCl的0V下测量过氧化氢。这种方法可显着减少干扰,尤其是易氧化的化合物(例如邻苯二酚和抗坏血酸)所产生的干扰。同样,在该工作电位下没有观察到明显的氧干扰。仅仔细研究了几种电解质。校准曲线在不超过1.3 mM(r = 0.9992)的条件下是线性的,注入0.05 mM过氧化氢时传感器的典型响应约为5.6 s。带有低通滤波器(时间常数0.1 s)的该化学传感器的检测极限为3.0 X 10(-8)M(S / N = 3),传感器的灵敏度为11.52 X 10(-6)A 1 mmol(-1)毫米(-2)(r = 0.9992)。该传感器的RSD为1.1%。 [参考:30]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号