首页> 外文期刊>Technology and health care: official journal of the European Society for Engineering and Medicine >Measurement and analysis of channel attenuation characteristics for an implantable galvanic coupling human-body communication
【24h】

Measurement and analysis of channel attenuation characteristics for an implantable galvanic coupling human-body communication

机译:植入式电流耦合人体通信的通道衰减特性的测量和分析

获取原文
获取原文并翻译 | 示例
           

摘要

In this study, an experiment was designed to verify the low power consumption of galvanic coupling human-body communication. A silver electrode (silver content: 99%) is placed in a pig leg and a sine wave signal with the power of 0 dBm is input. Compared with radio frequency communication and antenna transmission communication, attenuation is reduced by approximately 10 to 15 dB, so channel characteristics are highly improved.
机译:在这项研究中,设计了一个实验来验证电耦合人体通信的低功耗。将银电极(银含量:99%)放在猪脚中,并输入功率为0 dBm的正弦波信号。与射频通信和天线传输通信相比,衰减降低了大约10到15 dB,因此信道特性得到了极大的改善。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号