Based on a molecularly imprinted polymer (MIP) as the recognition element, a novel chemiluminescence (CL) micro flow sensor on a chip for the determination of terbutaline in human serum is described. The MIP was prepared by using terbutaline as the template, methacrylic acid (MAA) as the functional monomer, ethylene glycol dimethacrylate (EGDMA) as the cross-linking monomer, and acetonitrile as the solvent. The chip was fabricated from two 50 mm x 40 mm x 5 turn transparent poly (methylmethacrylate) (PMMA) slices. The microchannels on the chip etched by CO2 laser were 200 mu m wide and 150 mu m deep. The microsensor cell filled with 2 mg MIP for selectively on line adsorbing terbutaline was 10 mm long, 1 mm wide, and 0.5 mm deep. All reagents were controlled by the syringe pump with an accurate timer. The on line adsorbed terbutaline by the MIP can enhance the CL intensity of the reaction of luminol with ferricyanide. The enhanced CL intensity is linear with terbutaline concentration from 8.0 to 100 ng/mL with a detection limit of 4.0 ng/mL (3 sigma). The micro flow sensor provides for good reproducibility with the relative standard deviation of 3.6% (n = 7) for 20 ng/mL terbutaline. (c) 2006 Elsevier B.V. All rights reserved.
机译:基于分子印迹聚合物(MIP)作为识别元件,描述了一种新型化学发光(CL)微芯片传感器,用于测定人血清中的特布他林。 MIP的制备方法是使用特布他林作为模板,甲基丙烯酸(MAA)作为功能单体,乙二醇二甲基丙烯酸酯(EGDMA)作为交联单体,乙腈作为溶剂。该芯片由两个50毫米x 40毫米x 5圈透明聚(甲基丙烯酸甲酯)(PMMA)切片制成。用CO2激光蚀刻的芯片上的微通道宽200微米,深150微米。装有2 mg MIP的微传感器单元,用于选择性在线吸附特布他林,长10 mm,宽1 mm,深0.5 mm。所有试剂均由带有精确计时器的注射泵控制。 MIP在线吸附的特布他林可以提高鲁米诺与铁氰化物反应的CL强度。增强的CL强度与特布他林浓度为8.0至100 ng / mL呈线性关系,检出限为4.0 ng / mL(3 sigma)。微流量传感器具有20 ng / mL特布他林的相对标准偏差,为3.6%(n = 7),具有良好的重现性。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号