首页> 外文期刊>Talanta: The International Journal of Pure and Applied Analytical Chemistry >Speciation and preconcentration of inorganic antimony in waters by Duolite GT-73 microcolumn and determination by segmented flow injection-hydride generation atomic absorption spectrometry(SFI-HGAAS)
【24h】

Speciation and preconcentration of inorganic antimony in waters by Duolite GT-73 microcolumn and determination by segmented flow injection-hydride generation atomic absorption spectrometry(SFI-HGAAS)

机译:Duolite GT-73微柱对水中的锑进行形态分析和富集,并采用分段流动注射氢化物发生原子吸收光谱法(SFI-HGAAS)测定

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A selective matrix removal/separation/enrichment method,utilizing a microcolumn of a chelating resin with -SH functional groups(Duolite GT-73),was proposed for the determination of Sb(III)in waters by segmented flow injection-hydride generation atomic absorption spectrometry(SFI-HGAAS).The resin was selective to Sb(III)at almost all pH and acidity values employed,whereas Sb( V)was not retained at all and could be determined after a pre-reduction step with L-cysteine.Spike recoveries were tested at various concentration levels in different water types and were found to vary between 85 and 118%.Accuracy of the proposed methodology was checked by analyzing a standard reference material and a good correlation was found between the determined(13.3 +-1.1 mug l~(-1))and the certified value(13.79 +-0.42 mug l~(-1)).The method was applied to several bottled drinking water samples for antimony determination with and without preconcentration and none of the samples were found to contain antimony above the permissible level(5 mug l~(-1)).The characteristic concentration(the concentration of the analyte corresponding to an absorbance of 0.0044)was 0.55 mug l~(-1)and the 3 s limit of detection(LOD)based on five times preconcentration was 0.06 mug l~(-1).The applicability of the microcolumn separation/preconcentration/matrix removal method for flow injection systems was also demonstrated.
机译:提出了一种利用具有-SH官能团的螯合树脂微柱(Duolite GT-73)的选择性基体去除/分离/富集方法,通过分段流动注射氢化物发生原子吸收法测定水中的Sb(III)在几乎所有使用的pH和酸度值下,该树脂对Sb(III)都具有选择性,而Sb(V)根本没有保留,可以在用L-半胱氨酸进行预还原步骤后确定。在不同浓度的不同水类型下测试峰值回收率,发现回收率介于85%和118%之间。通过分析标准参考物质检查了所提出方法的准确性,并且确定的方法之间具有良好的相关性(13.3 + -1.1)杯l〜(-1))和标准值(13.79 + -0.42杯l〜(-1))。该方法适用于几种瓶装饮用水样品中的锑测定,无预浓缩,也没有发现含有锑高于允许水平(5杯l〜(-1))。特征浓度(对应于吸光度0.0044的分析物浓度)为0.55杯l〜(-1)和3 s的检测极限(LOD)基于五次预浓缩的结果是0.06杯l〜(-1)。还证明了微柱分离/预浓缩/基质去除方法在流动注射系统中的适用性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号