首页> 外文期刊>Talanta: The International Journal of Pure and Applied Analytical Chemistry >Preconcentration and separation of total mercury in environmental samples using chemically modified chloromethylated polystyrene-PAN (ion-exchanger) and its determination by cold vapour atomic absorption spectroemtry
【24h】

Preconcentration and separation of total mercury in environmental samples using chemically modified chloromethylated polystyrene-PAN (ion-exchanger) and its determination by cold vapour atomic absorption spectroemtry

机译:化学修饰的氯甲基化聚苯乙烯-PAN(离子交换剂)对环境样品中的总汞进行预浓缩和分离,并通过冷蒸气原子吸收光谱法测定

获取原文
获取原文并翻译 | 示例

摘要

The use of chemically modified chioromethylated polystyrene-PAN, CMPS-PAN (ion-exchanger) for the precon-centration and separation of total mercury after digestion in preparation for determination by cold vapour atomic absorption spectrometry (CVAAS) was described. The effects on the percentage of recovered mercury by mass chance of ion-exchanger, stirring time, pH of the solution samples and eluent concentration were studied. The distribution coefficient Kd is 106.6 ml g ~. The interfering effects of some foreign ions were described. The metal complex formed between CMPS-PAN ion-exchanger and mercury was characterized by IR spectroscopy, p1-1-metric titration and thermal analysis. The method is simple and rapidly applicable for the determination of total mercury (ng ml —natural water, milk and urine.
机译:描述了使用化学改性的氯甲基化聚苯乙烯-PAN,CMPS-PAN(离子交换剂)进行消解后对总汞的预浓缩和分离,以准备通过冷蒸气原子吸收光谱法(CVAAS)测定。研究了离子交换剂质量机会,搅拌时间,溶液样品的pH值和洗脱液浓度对汞回收率的影响。分配系数Kd为106.6ml g〜。描述了一些外来离子的干扰作用。 CMPS-PAN离子交换剂与汞之间形成的金属络合物通过红外光谱,p1-1滴定法和热分析进行了表征。该方法简便,快速,可用于测定总汞(ng毫升-天然水,牛奶和尿液)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号