首页> 外文期刊>Talanta: The International Journal of Pure and Applied Analytical Chemistry >Analysis of nitrosamines in water by automated SPE and isotope dilution GC/HRMS - Occurrence in the different steps of a drinking water treatment plant, and in chlorinated samples from a reservoir and a sewage treatment plant effluent
【24h】

Analysis of nitrosamines in water by automated SPE and isotope dilution GC/HRMS - Occurrence in the different steps of a drinking water treatment plant, and in chlorinated samples from a reservoir and a sewage treatment plant effluent

机译:通过自动SPE和同位素稀释GC / HRMS分析水中的亚硝胺-在饮用水处理厂的不同步骤中以及在水库和污水处理厂废水中的氯化样品中发生

获取原文
获取原文并翻译 | 示例
           

摘要

A method based on automated solid-phase extraction (SPE) and isotope dilution gas chromatography/high resolution mass spectrometry (GC/HRMS) has been developed for the analysis of nine nitrosamines in water samples. The combination of automated SPE and GC/HRMS for the analysis of nitrosamines has not been reported previously. The method shows as advantages the selectivity and sensitivity of GC/HRMS analysis and the high efficiency of automated SPE with coconut charcoal EPA 521 cartridges. Low method detection limits (MDLs) were achieved, along with a greater facility of the procedure and less dependance on the operator with regard to the methods based on manual SPE. Quality requirements for isotope dilution-based methods were accomplished for most analysed nitrosamines, regarding to trueness (80-120%), method precision (<15%) and MDLs (0.08-1.7 ng/L). Nineteen water samples (16 samples from a drinking water treatment plant {DWTP}, 2 chlorinated samples from a sewage treatment plant {STP} effluent, and 1 chlorinated sample from a reservoir) were analysed. Concentrations of nitrosamines in the STP effluent were 309.4 and 730.2 ng/L, being higher when higher doses of chlorine were applied. N-Nitrosodimethylamine (NOMA) and N-nitrosodiethylamine (NDEA) were the main compounds identified in the STP effluent, and NDEA was detected above 200 ng/L, regulatory level for NDMA in effluents stated in Ontario (Canada). Lower concentrations of nitrosamines were found in the reservoir (20.3 ng/L) and in the DWTP samples (n.d. -28.6 ng/L). NDMA and NDEA were respectively found in the reservoir and in treated and highly chlorinated DWTP samples at concentrations above 10 ng/L (guide value established in different countries). The highest concentrations of nitrosamines were found after chlorination and ozonation processes (ozonated, treated and highly chlorinated water) in DWTP samples. (C) 2008 Elsevier B.V. All rights reserved.
机译:已经开发了一种基于自动固相萃取(SPE)和同位素稀释气相色谱/高分辨率质谱(GC / HRMS)的方法来分析水中的9种亚硝胺。以前尚未报道过将自动SPE与GC / HRMS结合用于亚硝胺的分析。该方法的优势在于,GC / HRMS分析的选择性和灵敏度以及采用椰子木炭EPA 521柱的自动SPE的高效性。实现了较低的方法检出限(MDL),并且该程序的便利性更高,并且对基于手动SPE的方法的操作者的依赖性降低。对于大多数分析的亚硝胺,要达到基于同位素稀释法的质量要求,包括真实性(80-120%),方法精度(<15%)和MDL(0.08-1.7 ng / L)。分析了19个水样品(16个来自饮用水处理厂{DWTP}的样品,2个来自污水处理厂{STP}废水的氯化样品和1个来自水库的氯化样品)。 STP废水中亚硝胺的浓度为309.4和730.2 ng / L,当使用更高剂量的氯时,该浓度更高。 N-亚硝基二甲胺(NOMA)和N-亚硝基二乙胺(NDEA)是STP废水中鉴定出的主要化合物,安大略省(加拿大)中检测到的NDEA含量高于Nng监管水平200 ng / L。在储层和DWTP样品中(n.d. -28.6 ng / L)发现亚硝胺的浓度较低(20.3 ng / L)。 NDMA和NDEA分别在储层和经过处理的高氯化DWTP样品中发现,浓度超过10 ng / L(在不同国家/地区建立了指导值)。在DWTP样品中,经过氯化和臭氧化处理(臭氧化,处理和高度氯化的水)后,发现亚硝胺的浓度最高。 (C)2008 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号