A novel method for flow injection analysis of total antioxidant capacity using enzymatically produced ABTS(center dot+) and biamperometric detector containing interdigitated electrode

机译:酶法生产的ABTS(中心点+)和包含叉指电极的双安培检测器用于总抗氧化剂容量的流动注射分析的新方法

获取原文
获取原文并翻译 | 示例
           

摘要

Application of interdigitated array microelectrodes as electrochemical sensors for determination of antioxidant capacity is reported. Electrochemical measurements with interdigitated electrodes (IDE) were studied in both stationary solutions and the flow system. The method is based on biamperometric measurements using ABTS(center dot+)vertical bar ABTS redox couple in phosphate buffer solution, pH 7.40. During analysis, the ABTS radical cation was enzymatically produced by peroxidase in a tubular flow-through reactor. The performance of bioreactor was tested at different concentrations of immobilized enzyme, ABTS and hydrogen peroxide. The influence of flow rate on proper operation of the bioreactor was also studied. The results of antioxidant activity were determined using Trolox as a standard. The applied IDE detector accomplished good sensitivity of 0.3 nA/mu M of Trolox and offered linear range between 20 to 500 mu M of Trolox. The comparison of results (R-2 = 0.9915) for antioxidant activity between spectroscopic and FIA biamperometric measurements by interdigitated electrodes confirmed the applicability of the proposed method for determination of antioxidant capacity. (c) 2006 Elsevier B.V. All rights reserved.
机译:报道了叉指阵列微电极作为电化学传感器测定抗氧化能力的应用。在固定溶液和流动系统中都研究了采用叉指电极(IDE)进行的电化学测量。该方法基于双安培法测量,该测量是在pH 7.40的磷酸盐缓冲溶液中使用ABTS(中心点+)竖条ABTS氧化还原对进行的。在分析过程中,ABTS自由基阳离子是由管状过流反应器中的过氧化物酶酶促产生的。在不同浓度的固定化酶,ABTS和过氧化氢下测试了生物反应器的性能。还研究了流速对生物反应器正常运行的影响。抗氧化剂活性的结果以Trolox为标准测定。所应用的IDE检测器实现了0.3nA /μMTrolox的良好灵敏度,并提供了20至500μMTrolox的线性范围。通过叉指电极在光谱和FIA双安培测量之间的抗氧化剂活性结果(R-2 = 0.9915)的比较证实了所提出的抗氧化剂容量测定方法的适用性。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号