Microfluidic liquid-liquid extraction system based on stopped-flow technique and liquid core waveguide capillary

机译:基于停流技术和液芯波导毛细管的微流体液液萃取系统

获取原文
获取原文并翻译 | 示例
       

摘要

In this work, a miniaturized liquid-liquid extraction system under stopped-flow manipulation mode with spectrometric detection was developed. A Teflon AF liquid-core waveguide (LCW) capillary was used to serve as both extraction channel for organic solvent flow and adsorption detection flow cell. Gravity induced hydrostatic pressure was used to drive the organic and aqueous phases through the extraction channels. During extraction process, a stable organic and aqueous phase interface was formed at the outlet of the capillary, through which the analyte in the flowing aqueous stream was extracted into the stationary organic solvent in capillary. The absorbance of the analyte extracted into the organic solvent was measured in situ by a spectrometric detection system with light emitting diode (LED) as light source and photodiode as absorbance detector. The performance of the system was demonstrated in the determination of sodium dodecyl sulfate (SDS) extracted as an ion pair with methylene blue into chloroform. The precision of the measured absorbance for a 5 mg L-1 SDS standard was 6.1% R.S.D. (n = 5). A linear response range of 1-10 mg L-1 SDS was obtained with 5 min extraction period. The limit of detection (LOD) for SDS based on three times standard deviation of the blank response was 0.25 mg L-1. (c) 2006 Elsevier B.V. All rights reserved.
机译:在这项工作中,开发了一种在带光谱检测的停止流操作模式下的小型液-液萃取系统。使用Teflon AF液芯波导(LCW)毛细管作为有机溶剂流的提取通道和吸附检测流动池。重力感应的静水压力被用来驱动有机相和水相通过萃取通道。在萃取过程中,在毛细管的出口处形成了稳定的有机相和水相界面,通过该界面,流动的水流中的分析物被萃取到毛细管中的固定有机溶剂中。通过分光检测系统,以发光二极管(LED)为光源,光电二极管为吸光度检测器,对提取到有机溶剂中的分析物的吸光度进行原位测量。该系统的性能通过测定十二烷基硫酸钠(SDS)作为亚甲基蓝的离子对萃取到氯仿中来证明。 5 mg L-1 SDS标准品的吸光度测量精度为6.1%R.S.D. (n = 5)。在5分钟的萃取时间内,线性响应范围为1-10 mg L-1 SDS。基于空白响应三倍标准偏差的SDS检测限(LOD)为0.25 mg L-1。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号