A comparison of 500 prefilled textured saline breast implants versus 500 standard textured saline breast implants: is there a difference in deflation rates?

机译:比较500个预填充的有纹理的盐水乳房植入物与500个标准的有纹理的盐水乳房植入物:放气率是否有差异?

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: This study provides the first large-volume (1000 implant) comparison of the deflation rates of Poly Implant Prosthesis prefilled textured saline breast implants versus a control group of Mentor Siltex textured saline implants. METHODS: A consecutive series of 500 Poly Implant Prosthesis prefilled textured saline breast implants was compared with a consecutive series of 500 Mentor Siltex breast implants. Each breast implant was evaluated for a 4-year period, and the annual deflation rate (number of deflations during a given year divided by the total number of implants) and cumulative deflation rate (cumulative total of deflations through a given year divided by the total number of implants) were recorded. Statistical significance was calculated using the Fisher's exact test at year 1 and the chi-square analysis at years 2 through 4. RESULTS: The cumulative deflation rates of the Poly Implant Prosthesis implants was as follows: year 1, 1.2 percent; year 2, 5.6 percent; year 3, 11.4 percent; and year 4, 15.4 percent. The cumulative deflation rates of the Mentor implants was: year 1, 0.2 percent; year 2, 0.6 percent; year 3, 1.6 percent; and year 4, 4.4 percent. At year 1, the difference between deflation rates was not statistically significant (Fisher's exact test, p > 0.05). However, at year 2 (chi-square, 13.29; p < 0.001), year 3 (chi-square, 37.91; p < 0.001), and year 4 (chi-square, 32.69; p < 0.001), the difference was statistically significant. CONCLUSIONS: There was a statistically significant difference between the overall deflation rates of Poly Implant Prosthesis prefilled textured saline breast implants and Mentor Siltex breast implants at year 2, year 3, and year 4. After 4 years, the 15.56 percent cumulative deflation rate of Poly Implant Prosthesis implants was over 3.5 times higher than the 4.31 percent deflation rate of the Mentor Siltex implants. There may be several factors contributing to the higher deflation rate seen in Poly Implant Prosthesis implants, including possible invitro deflation before implantation and silicone shell curing technique. Nevertheless, this statistically significant deflation difference must be taken into account when balancing the risks and benefits of Poly Implant Prosthesis breast implants.
机译:背景:这项研究提供了第一个大批量(1000个植入物)比较聚植入物假体预填充纹理盐水乳房植入物与Mentor Siltex纹理盐水植入物对照组的放气率的方法。方法:将连续系列的500枚Poly植入物假体预填充纹理盐水乳房植入物与连续系列的500颗Mentor Siltex乳房植入物进行比较。对每个乳房植入物进行了为期4年的评估,年放气率(给定年份的放气次数除以植入物总数)和累积放气率(给定年份的总放气量除以总数)记录植入物的数量。使用第一年的Fisher精确检验和第二年至第四年的卡方分析计算统计显着性。结果:聚植入假体植入物的累积放气率如下:第一年为1.2%,第二年为1.2%。第二年,5.6%;第三年,11.4%;和第四年的15.4%。 Mentor植入物的累积放气率是:第一年为0.2%;第二年为0.2%。第二年,0.6%;第三年1.6%;和第四年,4.4%。在第一年,通货紧缩率之间的差异无统计学意义(Fisher精确检验,p> 0.05)。但是,在第2年(卡方,13.29; p <0.001),第3年(卡方,37.91; p <0.001)和第4年(卡方,32.69; p <0.001),差异具有统计学意义重大。结论:在第2年,第3年和第4年,Poly种植体假体预填充的生理盐水乳房植入物和Mentor Siltex乳房植入物的总体放气率在统计学上有显着差异。4年后,Poly累积放气率达到15.56%。植入物假体植入物比Mentor Siltex植入物的4.31%的放气率高3.5倍以上。在Poly Implant Prosthesis植入物中,可能有多种因素导致较高的放气速度,包括在植入前可能的体外放气和硅树脂外壳固化技术。但是,在平衡Poly Implant Prosthesis乳房植入物的风险和收益时,必须考虑到统计学上显着的通缩差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号