Determination of Ureter Stent Appearance on Dual-energy Computed Tomography Scan

机译:双能计算机断层扫描技术确定输尿管支架外观

获取原文
获取原文并翻译 | 示例
       

摘要

To examine the dual-energy computed tomography (DECT) properties of 7 commonly used ureteral jfents to optimize stent selection for calculi monitored using DECT. The use of DECT to evaluate renal and ureteral calculi has recently increased.Seven stents were individually placed in a fish bowl phantom and imaged using a Siemens Somatom Definition Flash CT scanner. DECT peak tube potentials of 80 and 140 kVp and 100 and 140 kVp were^used, reflecting our current dual-energy protocols. These were compared to 31 in vivo stents of mown composition. The data were reconstructed on a multimodality Work-Place (Siemens) using CT syngo Post-Processing Suite software.The average patient age was 64 years (range 27-90). The average body mass index was 31.9 kg/m2 (range 24-51.6). Of the 27 patients, 4 had uric acid stones and 22 had calcium-based stones; 1 patient had undergone renal transplantation. No difference was seen in the dual-energy characterization of stents from the same manufacturer. All imaged Cook and Bard stents had a dual-energy characterization that approached that of calcium stones (blue). All Boston Scientific and Gyms ACMI stents had a dual-energy characterization resembling that of uric acid stones (red).The present study evaluated the stent appearance on DECT for various stent manufacturers. This information will aid in the optimal stent selection for patients undergoing treatment of renal calculi and followed up with DECT.
机译:为了检查7种常用输尿管裂孔的双能计算机断层扫描(DECT)特性,以优化使用DECT监测的结石的支架选择。 DECT用于评估肾和输尿管结石的用途最近有所增加。将七个支架分别放置在鱼缸体模中,并使用Siemens Somatom Definition Flash CT扫描仪成像。使用了80和140 kVp以及100和140 kVp的DECT峰值管电势,反映了我们当前的双能方案。将这些与31个已割组合物的体内支架进行比较。使用CT syngo Post-Processing Suite软件在多模式工作场所(Siemens)上重建数据,平均患者年龄为64岁(范围27-90)。平均体重指数为31.9 kg / m2(范围24-51.6)。在这27例患者中,有4例有尿酸结石,有22例有钙基结石。 1例患者接受了肾脏移植。在同一制造商的支架的双能表征中未见差异。所有成像的Cook和Bard支架的双重能量特征都接近钙结石(蓝色)。所有Boston Scientific and Gyms ACMI支架均具有类似于尿酸结石(红色)的双能特性。本研究评估了DECT上各种支架制造商的支架外观。这些信息将有助于为接受肾结石治疗并接受DECT随访的患者选择最佳的支架。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号