Dual source computed tomography: a novel technique to determine stone composition.

机译:双源计算机断层扫描:一种确定石头成分的新技术。

获取原文
获取原文并翻译 | 示例
       

摘要

INTRODUCTION: Dual source computed tomography (CT), a novel technology that employs two different x-ray sources, may provide an image resolution that has not been possible with conventional single source CT. We tested the ability of dual source CT imaging to distinguish calcium oxalate (CaOx) stones from calcium phosphate (CaP) stones, and both types of calciferous stones from uric acid (UA) stones. METHODS: CaOx, CaP, and UA stones were placed in a tissue phantom and imaged with dual source multidetector CT at both 80kV and 140 kV. Hounsfield units (HU) of each stone were recorded for the 140 kV and the 80 kV datasets by drawing regions of interest by circle, hand-drawing, and using a volume calculation method. The differences between, and the ratios of, the HU obtained by the two energy sources were compared among the stone groups. RESULTS: Between CaOx and CaP stones, HU differences (80kV-140kV difference, hand-drawing technique: CaOx = 270.7; CaP = 369.5; UA = 11.45; P = 0.020 for CaOx vs. CaP; P <0.001 for CaOx and CaP vs. UA) as well as ratios (140kV:80kV ratio, hand-drawing technique: CaOx = 1.44; CaP = 1.51; UA = 1.04; P = 0.001 for CaOx vs. CaP; P < 0.001 for CaOx and CaP vs. UA) were significantly different. There were no significant differences in stone size among the groups. CONCLUSIONS: Dual source CT has the ability to distinguish not only between calciferous stones and uric acid stones, but also among stones composed of different calcium salts. Further studies are warranted to confirm these findings in a clinical setting.
机译:简介:双源计算机断层扫描(CT)是一种采用两个不同的X射线源的新颖技术,它可以提供传统的单源CT无法实现的图像分辨率。我们测试了双源CT成像的能力,以区分草酸钙(CaOx)结石和磷酸钙(CaP)结石,以及两种类型的钙化结石和尿酸(UA)结石。方法:将CaOx,CaP和UA结石放置在组织体模中,并用双源多探测器CT在80kV和140kV电压下成像。通过圆形绘制,手工绘制并使用体积计算方法,记录140 kV和80 kV数据集的每块石头的Hounsfield单位(HU)。比较了两种能源在石块组中两种能源获得的HU之间的差异和比率。结果:在CaOx和CaP结石之间,HU差异(80kV-140kV差异,手绘技术:CaOx = 270.7; CaP = 369.5; UA = 11.45; P = 0.020 for CaOx vs.CaP; P <0.001 for CaOx和CaP vs 。UA)和比率(140kV:80kV比率,手绘技术:CaOx = 1.44; CaP = 1.51; UA = 1.04; CaOx vs. CaP的P = 0.001; CaOx和CaP vs UA的P <0.001)有很大的不同。各组之间的结石大小没有显着差异。结论:双源CT不仅可以区分钙质结石和尿酸结石,还可以区分由不同钙盐组成的结石。有必要做进一步的研究以在临床上证实这些发现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号