Evaluation of efficacy of novel optically activated digital endoscope protection system against laser energy damage.

机译:新型光学激活数字内窥镜保护系统对激光能量损坏的功效评估。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To evaluate the reliability and efficacy of a novel endoscope protection system (EPS) against direct laser energy damage. METHODS: We performed in vitro evaluations of a novel EPS prototype that uses optical feedback from the digital sensor of the DUR-D ureteroscope to terminate the laser energy on retraction of the laser fiber into the ureteroscope. We evaluated various speeds of retraction (0.5, 2.0, and 5.0 cm/s) in normal saline and various concentrations of indigo carmine and human blood. We also evaluated the protrusion distance at which shutdown occurred with the laser fiber cladding cleaved at 0, 3, and 5 mm from the end of the fiber. Twenty trials of each condition were performed. RESULTS: In normal saline and blood dilutions of up to 10 g/L, the EPS worked with 100% efficiency for all trials. For blood dilutions of 10 g/L or greater and indigo carmine concentrations of 0.16 g/L or greater, the reliability of the EPS deteriorated. Lasers stripped of 0, 3, and 5 mm of cladding initiated shutdown at 2.9 +/- 0.13, 5.1 +/- 0.09, and 8.2 +/- 0.15 mm from the ureteroscope distal end, respectively (P 0.01). A single DUR-D ureteroscope was used for all trials and remained completely intact through 120 retractions of the active laser fiber into the channel. CONCLUSIONS: In this evaluation, the novel EPS was highly effective and reliable. When using indigo carmine or high-density blood concentrations, the efficacy of the EPS was compromised. The EPS should be used to complement standard safe laser technique rather than to replace it.
机译:目的:评估新型内窥镜保护系统(EPS)防止直接激光能量损坏的可靠性和功效。方法:我们对新型EPS原型进行了体外评估,该原型使用来自DUR-D输尿管镜数字传感器的光反馈,在激光束缩回输尿管镜时终止激光能量。我们评估了在生理盐水和不同浓度的靛蓝胭脂红和人体血液中的各种回缩速度(0.5、2.0和5.0 cm / s)。我们还评估了当激光光纤包层从光纤末端切开0、3和5 mm时,关闭发生的突出距离。每个条件进行了二十次试验。结果:在生理盐水和血液稀释液中浓度高达10 g / L的情况下,EPS在所有试验中的效率均为100%。对于10 g / L或更高的血液稀释液和0.16 g / L或更高的靛蓝胭脂红浓度,EPS的可靠性下降。剥去0、3和5毫米包层的激光分别在距输尿管镜远端2.9 +/- 0.13毫米,5.1 +/- 0.09毫米和8.2 +/- 0.15毫米处启动关机(P <0.01)。所有试验均使用一台DUR-D输尿管镜,并通过将活动激光光纤缩回120通道而完全完好无损。结论:在该评估中,新型EPS是高效且可靠的。当使用靛蓝胭脂红或高密度血液浓度时,EPS的功效受到损害。 EPS应该用于补充标准安全激光技术,而不是替代它。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号