The application of Spatially-Modulated Ultrasound Radiation Force (SMURF) to shear modulus imaging is demonstrated in tissue-mimicking phantoms and porcine liver. Scanning and data acquisition was performed with a Siemens Antares ultrasound scanner and VF7-3 linear array operating at 4.21 MHz. Modulus estimates in uniform phantoms of Zerdine with shear moduli of 5.1 and 12.4 kPa exhibited standard deviations within 6% of the mean value. Zerdine spheres 1 cm in diameter (nominally 2.7, 4.7 and 15 kPa) in a 8 kPa (nominal) background are clearly resolved. Cross sectional images of a soft conical inclusion in a gelatin-based phantom indicate a spatial resolution of approximately 2.5 mm. Images of the shear modulus of an ex-vivo sample of porcine liver tissue show an average value of 3 kPa. A stifflesion induced with 0.5 mL of 10% glutaraldehyde is clearly visible as a region of shear modulus in excess of 10 kPa. A modulus gradient associated with the diffusion of the glutaraldehyde is visible. Two pulse sequences were examined, differing only in the timing of the beams used to generate the shear waves. Details of the beam sequences and subsequent signal processing are presented.
机译:在模仿组织的模型和猪肝脏中证明了空间调制超声辐射力(SMURF)在剪切模量成像中的应用。使用西门子Antares超声扫描仪和以4.21 MHz运行的VF7-3线性阵列进行扫描和数据采集。 Zerdine的统一模型中的模量估计值具有5.1和12.4 kPa的剪切模量,其标准偏差在平均值的6%之内。直径为1 cm(标称值为2.7、4.7和15 kPa)的Zerdine球在8 kPa(标称)背景下得到清晰解析。基于明胶的幻像中的软圆锥形夹杂物的横截面图像显示约2.5 mm的空间分辨率。猪肝组织的离体样品的剪切模量图像显示平均值为3 kPa。在超过10 kPa的剪切模量区域内,可以清楚地看到由0.5 mL的10%戊二醛引起的僵硬。可见与戊二醛扩散相关的模量梯度。检查了两个脉冲序列,仅在用于产生切变波的光束时序上有所不同。给出了波束序列和后续信号处理的详细信息。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号