High resolution ultrasound and magnetic resonance imaging of the optic nerve and the optic nerve sheath: anatomic correlation and clinical importance.

机译:视神经和视神经鞘的高分辨率超声和磁共振成像:解剖相关性和临床重要性。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: We performed a cadaver study to evaluate the accuracy of measurements of the optic nerve and the optic nerve sheath for high resolution US (HRUS) and magnetic resonance imaging (MRI). MATERIALS AND METHODS: Five Thiel-fixated cadaver specimens of the optic nerve were examined with HRUS and MRI. Measurements of the optic nerve and the ONSD were performed before and after the filling of the optic nerve sheath with saline solution. Statistical analysis included the calculation of the agreement of measurements and the evaluation of the intraobserver and interobserver variation. RESULTS: Overall a good correlation of measurement values between HRUS and MRI can be found (mean difference: 0.02-0.97 mm). The repeatability coefficient (RC) and concordance correlation coefficient (CCC) values were good to excellent for most acquisitions (RC 0.2-1.11 mm; CCC 0.684-0.949). The highest variation of measurement values was found for transbulbar sonography (RC 0.58-1.83 mm; CCC 0.615/0.608). CONCLUSION: If decisive anatomic structures are clearly depicted and the measuring points are set correctly, there is a good correlation between HRUS and MRI measurements of the optic nerve and the ONSD even on transbulbar sonography. As most of the standard and cut-off values that have been published for ultrasound are significantly lower than the results obtained with MRI, a reevaluation of sonographic ONSD measurement with correlation to MRI is necessary.
机译:目的:我们进行了尸体研究,以评估高分辨率US(HRUS)和磁共振成像(MRI)的视神经和视神经鞘的测量准确性。材料与方法:用HRUS和MRI检查5头Thiel固定的视神经尸体标本。在用盐溶液填充视神经鞘管之前和之后,进行视神经和ONSD的测量。统计分析包括测量一致性的计算以及观察者内和观察者间变异的评估。结果:总体上可以发现HRUS与MRI之间的测量值具有良好的相关性(平均差异:0.02-0.97 mm)。对于大多数采集(RC 0.2-1.11 mm; CCC 0.684-0.949),重复性系数(RC)和一致性相关系数(CCC)值的优劣。经皮超声检查发现测量值变​​化最大(RC 0.58-1.83 mm; CCC 0.615 / 0.608)。结论:如果清楚地描述了决定性的解剖结构并正确设置了测量点,则即使在经皮超声检查下,HRUS和MRI对视神经和ONSD的测量也具有良好的相关性。由于已发布的有关超声的大多数标准值和临界值均显着低于MRI获得的结果,因此有必要重新评估与MRI相关的超声ONSD测量值。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号