Echo-Tracking Method Improves the Reliability in Flow-Mediated Dilation Assessment by Measuring Brachial Artery Reactivity

机译:回声跟踪方法通过测量肱动脉反应性提高了血流介导的扩张评估的可靠性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The purpose of the study was to compare the intraobserver and interobserver reliability and agreement for measurement of flow-mediated dilation (FMD) between B-mode method and echo-tracking (ET) method. Twenty healthy volunteers (mean age, 31.6 +/- 9.2 years) underwent ultrasound examination by both B-mode and ET methods. Baseline brachial artery diameter, post-cuff release diameter, and FMD percent were assessed by each sonologist on 2 consecutive days. Reliability was assessed by intraclass correlation coefficients, and Bland-Altman plots were used to visually compare measurement bias and agreement by the 2 ultrasound methods. A total of 40 pairs of data were available for analysis. Excellent intraobserver and interobserver reliability values were found for all variables assessed by the 2 methods. The intraclass correlation coefficient values were higher for ET in both intraobserver and interobserver evaluations, but only for interobserver evaluations for post-cuff release diameter and FMD was there no overlap in the 95% confidence interval. The Bland-Altman plots showed that in 95% of the measurements, the percentage difference between ET and B-mode ultrasound techniques was within 18.1%, 19.4%, and 17.3% for baseline brachial artery diameter, post-cuff release diameter, and FMD percent, respectively. The results suggest that ET and B-mode methods are reproducible in assessing the FMD. The ET method improves the reliability of FMD assessment, but we cannot determine which measurement is better for FMD.
机译:本研究的目的是比较观察者之间和观察者之间的可靠性以及在B模式方法和回声跟踪(ET)方法之间测量流介导的扩张(FMD)的一致性。 20名健康志愿者(平均年龄31.6 +/- 9.2岁)接受了B型和ET方法的超声检查。每位超声医师在连续2天评估基线肱动脉直径,袖带释放直径和FMD百分比。通过组内相关系数评估可靠性,并使用Bland-Altman图通过两种超声方法直观比较测量偏差和一致性。共有40对数据可供分析。对于通过这两种方法评估的所有变量,发现观察者间和观察者间的可靠性极佳。在观察者内和观察者间评估中,ET的类内相关系数值均较高,但仅对于观察者间评估的袖带后释放直径和FMD,在95%置信区间内没有重叠。 Bland-Altman图显示,在95%的测量结果中,ET和B型超声技术之间的基线臂肱动脉直径,袖带释放直径和FMD的百分比差异在18.1%,19.4%和17.3%之内百分比。结果表明,ET和B模式方法在评估FMD方面具有可重现性。 ET方法提高了FMD评估的可靠性,但我们无法确定哪种测量方法更适合FMD。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号