Recent studies have shown the feasibility of tissue and contrast harmonic imaging with a prototype nonlinear intravascular ultrasound (IVUS) system using a conventional single-element rotating IVUS catheter. In this study, a dual-frequency transducer element was mounted in an IVUS catheter and its second harmonic imaging performance was investigated and compared with that of a conventional IVUS catheter. Hydrophone measurements showed a transmit efficiency improvement of >6 dB for the dual-frequency catheter at 20 MHz. In vitro phantom experiments showed a signal-to noise ratio improvement of >5 dB in second harmonic mode at 40 MHz (H40) with the dual-frequency catheter, when using equal transmit voltage for both catheters. Finally, in vivo experiments were conducted and showed image improvement in H40 acquisitions with respect to the conventional IVUS catheter.
机译:最近的研究表明,使用常规的单元件旋转IVUS导管使用原型非线性血管内超声(IVUS)系统进行组织和对比谐波成像的可行性。在这项研究中,将双频换能器元件安装在IVUS导管中,并对其二次谐波成像性能进行了研究,并将其与常规IVUS导管进行了比较。水听器的测量结果表明,双频导管在20 MHz时的发射效率提高了> 6 dB。体外体模实验表明,当两个导管使用相等的传输电压时,在双频模式下,在40 MHz(H40)的二次谐波模式下,信噪比提高了> 5 dB。最后,进行了体内实验,并显示了相对于常规IVUS导管在H40采集中的图像改善。