首页> 外文期刊>Quimica nova >INTERNAL STANDARD VERSUS EXTERNAL STANDARD CALIBRATION: AN UNCERTAINTY CASE STUDY OF A LIQUID CHROMATOGRAPHY ANALYSIS
【24h】

INTERNAL STANDARD VERSUS EXTERNAL STANDARD CALIBRATION: AN UNCERTAINTY CASE STUDY OF A LIQUID CHROMATOGRAPHY ANALYSIS

机译:内部标准与外部标准校准:液相色谱分析的不确定性案例研究

获取原文
获取原文并翻译 | 示例

摘要

Traditionally, in the cigarettes industry, the determination of ammonium ion in the mainstream smoke is performed by ion chromatography. This work studies this determination and compares the results of this technique with the use of external and internal standard calibration. A reference cigarette sample presented measurement uncertainty of 2.0 μg/cigarette and 1.5 μg/cigarette, with external and internal standard, respectively. It is observed that the greatest source of uncertainty is the bias correction factor and that it is even more significant when using external standard, confirming thus the importance of internal standardization for this correction.
机译:传统上,在卷烟工业中,主流烟气中的铵离子的测定是通过离子色谱法进行的。这项工作研究了这一确定结果,并将此技术的结果与使用外部和内部标准校准进行了比较。参比香烟样品的测量不确定度分别为2.0μg/支和1.5μg/支(内标和内标)。可以看出,不确定性的最大来源是偏差校正因子,当使用外部标准时,它甚至更为重要,因此确认了内部标准化对于这种校正的重要性。