Standards, calibration, and guidelines in microcalorimetry - Part 2. Calibration standards for differential scanning calorimetry - (IUPAC Technical Report)

机译:微量量热法的标准,校准和指南-第2部分。差示扫描量热法的校准标准-(IUPAC技术报告)

获取原文
获取原文并翻译 | 示例
       

摘要

Differential scanning calorimeters (DSCs) are widely used for temperature, heat capacity, and enthalpy measurements in the range from subambient to high temperatures. The present recommendations describe procedures and reference materials (RMs) for the calibration of DSCs. The recommendations focus on the calibration of the response of the instrument and on the estimation of the measurement uncertainty. The procedures for temperature, enthalpy, and heat-flow rate calibration are given in detail. Calibration on cooling has also been considered. Recommended RMs are listed, and the relevant properties of these materials are discussed.
机译:差示扫描量热仪(DSC)广泛用于温度,热容量和焓的测量,范围从环境温度到高温。本建议介绍了DSC校准的程序和参考材料(RM)。这些建议集中在仪器响应的校准和测量不确定度的估计上。详细给出了温度,焓和热流率校准的程序。还考虑了冷却校准。列出了推荐的RM,并讨论了这些材料的相关特性。