Influence of contaminants in the stem-ball interface on the static fracture load of ceramic hip joint ball heads

机译:阀杆-球界面中的污染物对陶瓷髋关节球头静态断裂载荷的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The probability of in-vivo failure of ceramic hip joint implants is very low (0.05-0.004 per cent). Besides material flaws and overloading, improper handling during implantation may induce fractures of the ceramic ball head in the long term. This study focuses on the influence of contaminants located in the stem-ball interface and on the use of damaged metal tapers on the strength of ceramic ball heads. Mechanical tests on alumina ball heads according to the standard ISO 7206-10 were performed to identify their effect on the static fracture load. A decrease of up to 90 per cent with respect to the reference static fracture load was found when contaminants such as bone chips, soft tissue, or blood were present. Reductions of 57 per cent and 27 per cent were observed for deformed stem cross-sections (from circular to elliptical) and for flattened stems respectively, making deformed stems another influential parameter. Since any alteration of the interface between the metal taper and the ceramic ball head yields a nonuniform load introduction and hence results in stress concentrations, its presence has to be avoided.
机译:陶瓷髋关节植入物的体内失败可能性非常低(0.05-0.004%)。除了材料缺陷和过载外,长期植入还可能导致陶瓷球头破裂。这项研究的重点是位于球杆接口处的污染物的影响,以及使用受损的金属锥度对陶瓷球头的强度。根据标准ISO 7206-10对氧化铝球头进行了机械测试,以确认其对静态断裂载荷的影响。当存在诸如碎骨,软组织或血液之类的污染物时,相对于参考静态断裂载荷,降低了多达90%。变形茎的横截面(从圆形到椭圆形)和扁平茎分别减少了57%和27%,这使变形茎成为另一个有影响的参数。由于金属锥度和陶瓷球头之间的界面发生任何变化都会导致载荷引入不均匀,从而导致应力集中,因此必须避免其存在。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号