Shortening of an anatomical stem, how short is short enough? An in vitro study of load transfer and primary stability

机译:解剖茎短,足够短足够短吗?载荷传递和初级稳定性的体外研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An anatomical stem should be short enough to avoid distal locking and distal load transfer but long enough to ensure adequate primary stability of the stem. In this in vitro study, the cortical strains in the femur and the primary stability of the stem were measured after insertion of Anatomic Benoist Girard-I anatomical stems with gradually reduced stem length in six human cadaver femurs in order to find the optimal stem length. A shortening of 40-50 mm, corresponding to a stem extending 30-40mm below the lesser trochanter, did not affect stem stability but nearly normalized the load distribution in the lower metaphysis and upper diaphysis. The large strain shielding observed in the calcar region was not influenced by shortening of the stem.
机译:解剖柄应足够短以避免远端锁定和远端负荷转移,但又要足够长以确保柄具有足够的基本稳定性。在这项体外研究中,在插入六个人尸体股骨的股骨长度逐渐减小的解剖Benoist Girard-I解剖茎之后,测量股骨的皮质应变和股骨的初级稳定性,以找​​到最佳的股骨长度。缩短40至50毫米,对应于在小转子下方延伸30至40毫米的茎,并没有影响茎的稳定性,但几乎使下干physi端和上干nearly端的负荷分布正常化。在足骨区域观察到的大应变屏蔽不受茎杆缩短的影响。