首页> 外文期刊>Physiotherapy theory and practice >A new clinical device for measuring wrist strength in people with tetraplegia.
【24h】

A new clinical device for measuring wrist strength in people with tetraplegia.

机译:一种用于测量四肢瘫痪患者腕力的新型临床设备。

获取原文
获取原文并翻译 | 示例
           

摘要

Weakness of wrist muscles has a significant impact on upper limb function in people with tetraplegia. A simple device has been designed to enable clinicians to measure dynamic strength in the very weak muscles of people with tetraplegia. The objective of this study was to test the interrater reliability of the device. Twenty wrists of 12 people with tetraplegia and associated wrist extensor weakness (grade 2-4) were included in the study. The strength of the wrist extensor muscles was measured by blinded assessors on two separate occasions 1-3 days apart. Measurements were conducted in a device that ensured the resistive torque was constant as the wrist moved through full range of motion. Intraclass correlation coefficients and percent close agreement scores were used to measure reliability. The intraclass correlation coefficient was 0.88 (95% CI 0.72-0.95). Repeat measurements of wrist extensor torque were within 0.5 Nm 75% of the time and 0.7 Nm 90% of the time. This device is a simple and reliable way of measuring dynamic wrist extensor strength of people with tetraplegia. It is appropriate for use in the clinical setting.
机译:腕肌无力对四肢瘫痪患者的上肢功能有重要影响。设计了一种简单的设备,使临床医生能够测量四肢瘫痪患者非常弱的肌肉中的动态力量。这项研究的目的是测试设备的可靠性。该研究包括了12例四肢瘫痪及相关腕伸肌无力(2-4级)的人的20个手腕。盲法评估者在相隔1-3天的两个不同场合测量腕部伸肌的力量。在确保手腕在整个运动范围内移动时阻力扭矩恒定的设备中进行测量。类内相关系数和紧密一致性百分比分数用于衡量可靠性。组内相关系数为0.88(95%CI 0.72-0.95)。重复测量腕部伸肌扭矩的时间在0.5 Nm的75%以内,在0.7 Nm的90%内。该设备是测量四肢瘫痪者动态腕部伸肌力量的简单可靠的方法。适合在临床环境中使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号