Home blood pressure (BP) monitoring is routinely recommended to hypertensive patients in the ambulatory care setting. Patients may choose from a variety of monitors, but the accuracy of these devices may be variable, and store brands are rarely included in comparison studies of these monitors. This study compared BP measurements obtained with the Omron Auto-Inflate Blood Pressure Monitor with Intellisense (HEM-711AC) and the ReliOn Auto-Inflate Digital Blood Pressure Monitor (HEM-741CREL) with a standard of manual readings obtained by auscultation with a stethoscope and aneroid sphygmomanometer. Participants were patients seeing the clinical pharmacist at an ambulatory clinic affiliated with a major teaching hospital and who consented to having their BPs measured using each of the 3 methods. Whereas both automatic monitors tended to overestimate systolic BP (SBP) and underestimate diastolic BP compared with the manual method, the only significant difference between the 3 devices was the Omron monitor's 2.27-mm Hg average overestimation of SBP (95% confidence interval, 0.62-3.92 mm Hg, P = 0.007). The ReliOn monitor did not differ from manual method any more or less than did the Omron monitor in terms of SBP (Wilcoxon signed rank test, P = 0.4512), diastolic BP (P = 0.1398), or pulse (P = 0.9718). Results were robust throughout subgroup analyses for different BP levels and order of device use. Either home monitor included in this study could be recommended for patient use to compliment standard office BP readings.
机译:通常建议在门诊护理环境中对高血压患者进行家庭血压(BP)监测。患者可以从各种监视器中进行选择,但是这些设备的准确性可能会有所不同,并且商店品牌很少包含在这些监视器的比较研究中。这项研究将带Intellisense的Omron自动充气血压监测仪(HEM-711AC)和ReliOn自动充气数字血压监测仪(HEM-741CREL)获得的血压测量值与通过听诊器和听诊器听诊获得的手动读数标准进行了比较无液血压计。参加者是在主要教学医院附属的门诊就诊的临床药剂师,并同意使用这三种方法中的每一种方法测量血压的患者。与手动方法相比,两个自动监护仪都倾向于高估收缩压(SBP)和低估舒张压,这3种设备之间的唯一显着差异是欧姆龙监护仪对SBP的平均估测值为2.27 mm Hg(95%置信区间,0.62- 3.92毫米汞柱,P = 0.007)。 ReliOn监护仪在SBP(Wilcoxon有符号秩检验,P = 0.4512),舒张压BP(P = 0.1398)或脉搏(P = 0.9718)方面与Omron监护仪没有任何不同。在不同血压水平和器械使用顺序的整个亚组分析中,结果都是可靠的。可以建议本研究中包括的任何家用监护仪供患者使用,以补充标准的办公室BP读数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号