Evaluation of a New Rapid and Fully Automated Electrochemiluminescence Immunoassay for Thyrotropin Receptor Autoantibodies

机译:新型快速,全自动电化学发光免疫分析法测定促甲状腺素受体自身抗体的价值

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Hyperthyroidism in Graves' disease is caused by autoantibodies to the TSH receptor (TSHR), and measurement of the TSHR autoantibody (TRAb) yields important information to diagnose and decide on the course of treatment of Graves' disease. We evaluated basic and clinical performance of a new, rapid, and fully automated electrochemiluminescence immunoassay Elecsys Anti-TSHR (Elecsys TRAb) for measuring serum TRAb.nnMethods: For evaluation of basic performance of the assay, we carried out intra- and interassay precision studies using five serum pools and three serum pools, respectively, and the assay was compared with four commercial TRAb assays. Clinical performance of the assay was evaluated with sera from 298 patients with untreated Graves' disease, 220 patients with destructive (painless and subacute) thyroiditis, and 332 healthy volunteers. The optimal cutoff point, which was calculated by receiver operating characteristic (ROC) analysis with the above subjects, was then used to classify an independent sample set of 80 patients with untreated Graves' disease, and 152 patients with destructive thyroiditis.nnResults: Intraassay coefficient of variation (CV) was 4.24% at 1.85IU/L and interassay CV was 10.1% at 1.46IU/L. All the correlation coefficient values calculated against four commercial assays were larger than 0.85. ROC analysis resulted in a specificity of 99.1% with a sensitivity of 97.0% at a decision limit of 1.86IU/L from comparison with untreated Graves' disease and destructive thyroiditis. The cutoff point yielded a sensitivity of 87.5% and specificity of 96.7% with the independent sample set.nnConclusion: In spite of the short measuring time of only 27 minutes, the assay showed the same or better results with the existing commercial products. The short measuring time would contribute to speedy, preconsultation diagnosis of thyroid disease, especially of Graves' disease.
机译:背景:格雷夫斯病的甲状腺功能亢进症是由针对TSH受体(TSHR)的自身抗体引起的,对TSHR自身抗体(TRAb)的测量可为诊断和决定格雷夫斯病的治疗过程提供重要信息。我们评估了一种新的,快速且全自动的电化学发光免疫测定Elecsys Anti-TSHR(Elecsys TRAb)用于测量血清TRAb的基本和临床性能。分别使用五个血清池和三个血清池,并将该测定与四种商业TRAb测定进行比较。用来自298例未治疗Graves病的患者,220例具破坏性(无痛和亚急性)甲状腺炎的患者和332名健康志愿者的血清评估了该测定的临床性能。最佳临界点是通过上述受试者的受试者工作特征(ROC)分析计算得出的,然后被用于对80例未经治疗的Graves病患者和152例破坏性甲状腺炎患者的独立样本集进行分类。变异(CV)在1.85IU / L时为4.24%,批间CV在1.46IU / L时为10.1%。对照四种商业测定法计算的所有相关系数值均大于0.85。与未经治疗的Graves病和破坏性甲状腺炎相比,ROC分析得出的特异性为99.1%,灵敏度为97.0%,决策极限为1.86IU / L。截止点对独立样品产生的灵敏度为87.5%,特异性为96.7%。nn结论:尽管测量时间短,只有27分钟,但该检测结果与现有的商业产品相同或更好。较短的测量时间将有助于对甲状腺疾病,尤其是格雷夫斯病进行快速的咨询前诊断。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号