首页> 外文期刊>Radiological Physics and Technology >Simplified simulation of four-layer depth of interaction detector for PET
【24h】

Simplified simulation of four-layer depth of interaction detector for PET

机译:PET四层相互作用深度检测仪的简化模拟

获取原文
获取原文并翻译 | 示例

摘要

A four-layer depth-of-interaction (DOI) detector was proposed from and developed at the research project at the National Institute of Radiological Sciences in Japan, with the aim of achieving high resolution and high spatial sensitivity. Previously, we had developed a Monte Carlo simulation for a DOI detector with 2 × 2 × 4 crystal elements. In this study, in order to simulate the final version of the DOI detector of the project, which uses a larger number of crystal elements, we have developed a much faster and simpler simulator. In this paper the algorithm of the simplified simulator as well as the previously proposed Monte Carlo method is presented and the validation of the simplified simulator through comparisons with the full Monte Carlo simulation and with some experimental results is described.
机译:日本国立放射科学研究所的研究项目提出并开发了一种四层相互作用深度(DOI)检测器,旨在实现高分辨率和高空间灵敏度。以前,我们已经为具有2×2×4晶体元素的DOI检测器开发了Monte Carlo模拟。在本研究中,为了模拟使用大量晶体元素的该项目DOI检测器的最终版本,我们开发了一种更快,更简单的模拟器。本文介绍了简化模拟器的算法以及先前提出的蒙特卡洛方法,并通过与完整的蒙特卡洛模拟的比较和一些实验结果对简化模拟器的验证进行了描述。