A nationwide survey was conducted on electron accelerators for radiation therapy in Japan for obtaining fundamental data on the actual operational situation of these accelerators. The research subjects were all medical institutions equipped with electron accelerators for radiation therapy (874 accelerators) in Japan. Postcards explaining the purpose and procedure of the survey were sent to each institution. Each institution was requested to download a questionnaire from the website and return it by e-mail. We obtained the following results. Energy: about 90% of the accelerators did not exceed 10 MeV as the permitted maximum energy. Beam intensity: a total of 94% of the accelerators did not exceed 4 Gy/min as the mean exposure dose rate. Beam time: a total of 74% of the accelerators were below half of the permitted maximum beam time. However, 2% of the accelerators exceeded 90% of the permitted maximum time. Cumulative dose: about 10% of institutions obtained permission not to use the maximum beam time, but to use the maximum cumulative dose to secure sufficient beam time. From these results, it became clear that the induced radioactivity based on the maximum beam current and the maximum operating time under an official permit would be overestimated. Therefore, the neutron dose should be assessed by use of a daily operational record, not from an official permit.
机译:在日本,对用于放射治疗的电子加速器进行了全国性调查,以获取有关这些加速器实际运行状况的基本数据。研究对象是日本所有配备了用于放射治疗的电子加速器(874加速器)的医疗机构。解释调查目的和程序的明信片已寄给每个机构。要求每个机构从网站上下载一份调查表,并通过电子邮件将其退回。我们获得了以下结果。能量:约90%的加速器未超过10 MeV作为允许的最大能量。束强度:总共94%的促进剂的平均暴露剂量率不超过4 Gy / min。束流时间:总共74%的加速器低于允许的最大束流时间的一半。但是,有2%的加速器超过了允许的最大时间的90%。累积剂量:约有10%的机构获得许可,不使用最大射束时间,而是使用最大累积剂量以确保足够的射束时间。从这些结果可以明显看出,根据最大束流和官方许可下的最大工作时间所产生的放射性将被高估。因此,中子剂量应通过日常运行记录而不是从官方许可中评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号