首页> 外文期刊>Quality engineering >Calibration, Error Analysis, and Ongoing Measurement Process Monitoring for Mass Spectrometry
【24h】

Calibration, Error Analysis, and Ongoing Measurement Process Monitoring for Mass Spectrometry

机译:质谱的校准,误差分析和持续测量过程监控

获取原文
获取原文并翻译 | 示例
           

摘要

We consider problems of quantifying and monitoring accuracy and precision of measurement in mass spectrometry, particularly in contexts where there is unavoidable day-to-day/ period-to-period changes in instrument sensitivity. First, we consider the issue of estimating instrument sensitivity based on data from a typical calibration study. Simple method-of-moments methods, likelihood-based methods, and Bayes methods based on the one-way random effects model are illustrated. Then, we consider subsequently assessing the precision of an estimate of a mole fraction of a gas of interest in an unknown. Finally, we turn to the problem of ongoing measurement process monitoring and illustrate appropriate setup of Shewhart control charts in this application.
机译:我们考虑量化和监控质谱法中测量准确性和精确性的问题,尤其是在仪器灵敏度每天/每周期都有不可避免的变化的情况下。首先,我们考虑基于典型校准研究中的数据估算仪器灵敏度的问题。说明了基于单向随机效应模型的简单矩量法,基于似然法和贝叶斯方法。然后,我们考虑随后评估未知气体中目标气体的摩尔分数的估计精度。最后,我们转向正在进行的测量过程监控的问题,并说明此应用程序中Shewhart控制图的适当设置。