首页> 外文期刊>Powder diffraction >Improved bench-top X-ray diffractometer for crystalline phase analysis of cement
【24h】

Improved bench-top X-ray diffractometer for crystalline phase analysis of cement

机译:改进的台式X射线衍射仪,用于水泥的结晶相分析

获取原文
获取原文并翻译 | 示例
           

摘要

A laboratory-scale X-ray diffractometer for obtaining high X-ray intensity data was developed. The apparatus, equipped with 600 W CuKα radiation with a Ni filter, incorporates technology that dramatically improves the quality of X-ray diffraction. The system comprises of a low-noise one-dimensional silicon strip detector, a variable slit, and a goniometer with a radius as small as 150 mm. A variable knife edge was used as a countermeasure for unwanted scattering, particularly in the low angle range. With this system, cement may be analyzed within 5 min. NIST 2686 standard reference material was analyzed using the newly developed diffractometer, and the quantitative analysis results for the major phases are in agreement with the certified values of the reference material.
机译:开发了用于获得高X射线强度数据的实验室规模的X射线衍射仪。该仪器配备了带Ni滤光片的600 WCuKα辐射,并结合了可显着提高X射线衍射质量的技术。该系统包括一个低噪声的一维硅条检测器,一个可变缝和一个半径小至150 mm的测角仪。可变的刀刃被用作对付不必要散射的对策,尤其是在低角度范围内。使用该系统,可以在5分钟内分析水泥。使用最新开发的衍射仪对NIST 2686标准参比物进行了分析,主要相的定量分析结果与参比物的认证值相符。

著录项

  • 来源
    《Powder diffraction》 |2013年第4期|249-253|共5页
  • 作者单位

    Application Laboratories, Rigaku Corporation, 3-9-12, Matsubara-cho, Akishima-shi Tokyo, Japan;

    Application Laboratories, Rigaku Corporation, 3-9-12, Matsubara-cho, Akishima-shi Tokyo, Japan;

    Application Laboratories, Rigaku Corporation, 3-9-12, Matsubara-cho, Akishima-shi Tokyo, Japan;

  • 收录信息 美国《科学引文索引》(SCI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Bench-top X-ray diffractometer; cement; Rietveld refinement; calcification process;

    机译:台式X射线衍射仪;水泥;Rietveld精致;钙化过程;