DESIGN OF A TREATMENT ROOM FOR AN 18-MV LINAC

机译:18 MV直线加速器治疗室的设计

获取原文
获取原文并翻译 | 示例
       

摘要

This work studies the design of the treatment room for an 18-MV linac to ensure radiation protection and safety of hospital staff and patients. The walls' thickness, the door, and the maze were designed according to the National Council on Radiation Protection and Measurements Report 151 recommendations. The results of this work are contrasted with the Monte Carlo calculations performed with the MCNP5 code where dose equivalents due to neutrons and neutron spectra estimated at different points inside and outside the radiotherapy room verify that the shielding thicknesses obtained are enough to reduce the dose level permitted by Mexican regulation.
机译:这项工作研究了18-MV直线加速器治疗室的设计,以确保辐射防护和医院工作人员及患者的安全。壁的厚度,门和迷宫是根据美国国家辐射防护和测量委员会151报告的建议设计的。这项工作的结果与MCNP5代码进行的蒙特卡洛计算形成对照,在MCNP5代码中,由中子和放射线室内外不同点估计的中子和中子光谱引起的剂量当量验证了所获得的屏蔽厚度足以降低允许的剂量水平根据墨西哥法规。

著录项

  • 来源
    《Nuclear Technology》 |2011年第1期|p.93-104|共12页
  • 作者单位

    Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares Apartado Postal 336, Zacatecas, 98000, Mexico;

    Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares Apartado Postal 336, Zacatecas, 98000, Mexico;

    Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares Apartado Postal 336, Zacatecas, 98000, Mexico;

    Instituto Zacatecano del Tumor, Apartado Postal 294, Zacatecas, 98000, Mexico;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    monte carlo methods; shielding design; linear accelerator;

    机译:蒙特卡洛方法屏蔽设计;线性加速器;
  • 入库时间 2022-08-18 00:43:46

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号