首页> 外文期刊>Nuclear Technology >Computer-Aided Star Shot Analysis for Linac Quality Assurance Testing
【24h】

Computer-Aided Star Shot Analysis for Linac Quality Assurance Testing

机译:直线加速器质量保证测试的计算机辅助星图分析

获取原文
获取原文并翻译 | 示例
       

摘要

With the increase in the precision of treatments delivered using radiotherapy machines, there has been a corresponding rise in demand for quality assurance tests that can verify the accuracy of these machines. One common test, star shot analysis, evaluates the isocenter stability of a radiotherapy machine using radiosensitive film or the electronic portal imaging device (EPID). This work details the development of an in-house method of automatically processing film and EPID images to conduct quality assurance testing. In contrast to commercially available software that analyzes a composite image star shot with multiple spokes superimposed on a single image, this work investigates a Gaussian peak finding technique while leveraging the EPID to image one spoke at a time.Spoke-by-spoke analysis was used to investigate the effects of opposing angles on composite image star shot analysis and to assess for collimator trajectories with minimal walkout. This revealed that irradiating film using opposing angles can give artificially low variations in the radiation isocenter due to offsetting deviations from the true center and that walkout was not the same for every 180-deg arc for the collimator, implying that some rotation arcs could give less variation during treatment.
机译:随着使用放射治疗机进行治疗的精度的提高,对可以验证这些机器的准确性的质量保证测试的需求也相应增加。一项星光分析是一项常见的测试,它使用放射敏感性胶片或电子门成像设备(EPID)来评估放射治疗机的等中心稳定性。这项工作详细介绍了自动处理胶片和EPID图像以进行质量保证测试的内部方法的开发。与市售的软件可以分析在一张图像上叠加了多个辐条的合成图像星形照片相比,这项工作研究了高斯峰发现技术,同时利用EPID一次对一个辐条进行成像。研究对角对合成图像星图分析的影响,并以最小的走动来评估准直器的轨迹。这表明使用相对角度的辐照薄膜可以抵消与真实中心之间的偏差,从而使人为辐射等中心线的变化很小,并且准直仪的每180度弧线的走动都不相同,这意味着某些旋转弧可以减少治疗期间的变化。

著录项

  • 来源
    《Nuclear Technology》 |2019年第7期|905-911|共7页
  • 作者单位

    Georgia Inst Technol, Med Phys Dept, 770 State St, Atlanta, GA 30318 USA;

    Georgia Inst Technol, Med Phys Dept, 770 State St, Atlanta, GA 30318 USA|HCA Healthcare, 2410 Patterson St,Basement Level, Nashville, TN 37203 USA;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Star shot analysis; electronic portal imaging device; linac;

    机译:星光分析;电子门禁成像设备;直线加速器;
  • 入库时间 2022-08-18 04:20:04

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号