The effect of lead in bone densitometry

机译:铅在骨密度测定中的作用

获取原文
获取原文并翻译 | 示例
           

摘要

Dual energy X-ray absorptiometry (DXA) is presently considered the standard technique for diagnosis of osteoporosis. It has been suggested that the presence of lead interferes with the accurate measurement of bone mineral density (BMD) by DXA because of the increased attenuation and that an accurate measurement of BMD cannot be determined unless the patient's bone lead content of patients is known. We performed DXA measurements on plaster of Paris phantoms and a Hologic Spine phantom in combination with polyester resin doped with various concentrations of lead. At lead levels which correspond to bone concentrations in occupationally exposed individuals, the suggested increase in densitometric BMD was not detected. Numerical calculations show that the effect of the lead depends upon the two energies of the X-ray beam of a particular device. The discrepancy between the actual and the densitometric BMD increases linearly and is about 0.3% at 100 ppm. Such change cannot be detected by the Hologic QDR 4500A, the device used for this experiment.
机译:目前,双能X线骨密度仪(DXA)被认为是诊断骨质疏松症的标准技术。已经提出,由于衰减增加,铅的存在干扰了DXA对骨矿物质密度(BMD)的精确测量,并且除非知道患者的患者骨铅含量,否则无法确定BMD的精确测量。我们对巴黎幻影和Hologic Spine幻影的石膏与掺杂了不同浓度铅的聚酯树脂进行了DXA测量。在对应于职业暴露个体的骨浓度的铅水平下,未检测到建议的光密度BMD升高。数值计算表明,导线的影响取决于特定设备的X射线束的两种能量。实际BMD与光密度BMD之间的差异线性增加,在100 ppm时约为0.3%。 Hologic QDR 4500A(此实验所用的设备)无法检测到这种变化。