The A-711 high yield neutron generator and automated pneumatic transfer system for fast neutron activation analysis

机译:A-711高产率中子发生器和自动气动传输系统,用于快速中子活化分析

获取原文
获取原文并翻译 | 示例
           

摘要

To make fast neutron activation analysis (FNAA) of samples with short half-lives easier, Thermo Electron has updated and modernized its automatic pneumatic transfer system for activation laboratories. For example, with a separation of 10 m from the counting station and a transit speed of 15 m/s, oxygen can be analyzed with improved accuracy. The fast transit time is needed due to the short half-lives of ~(16)N and ~(19)O, 7 s and 27 s respectively, and oxygen-free polyethylene sample bottles are used to allow prompt counting and decrease background counts. Incorporating a dual-axis rotator at the irradiation station for sample and standard, the transfer system also incorporates stations for sample loading, disposal and counting as well as a station to incorporate a chemical neutron source such as Cf-252.
机译:为了简化半衰期短的样品的快速中子活化分析(FNAA),赛默飞世尔(Thermo Electron)已更新和现代化了用于活化实验室的自动气动传输系统。例如,与计数站的距离为10 m,传输速度为15 m / s,因此可以以更高的精度分析氧气。由于〜(16)N和〜(19)O的半衰期较短,分别需要7 s和27 s,因此需要快速的运输时间,并且使用无氧聚乙烯样品瓶来快速计数并减少背景计数。传输系统在辐照站上装有一个用于样品和标准液的双轴旋转器,该传输系统还包括了用于样品加载,处理和计数的站以及一个装有化学中子源(例如Cf-252)的站。