Mass-resolved retarding field energy analyzer and its measurement of ion energy distribution in helicon plasma

机译:质量分辨延迟场能量分析仪及其在螺旋等离子体中离子能量分布的测量

获取原文
获取原文并翻译 | 示例
       

摘要

A novel type of plasma analyzer, which combines five-grid ion energy analyzer and a quadrupole mass spectrometer (QMS), has been developed to simultaneously measure the energy distribution and the mass of ions in a helicon plasma. The ions of the plasma were collected and analyzed through an aperture at the center of the sample holder so as to minimize plasma disturbance. The mass/charge ratios (m/e) of the three types of distinguishable ions in an argon plasma, Ar2+, Ar+ and Ar-2(+), are 20, 40 and 80, respectively. Fixing the mass of Ar ions one at a time, their energy distributions (IED) are measured at rf power of 1000 W. The results show that the fairly broad energy distributions of different kinds of ions are rather different. Possible mechanisms are tentatively explained. (c) 2005 Elsevier B.V. All rights reserved.