Neutron-gamma mixed field measurements by means of MCP-TLD600 dosimeter pair

机译:通过MCP-TLD600剂量计对进行中子伽马混合场测量

获取原文
获取原文并翻译 | 示例

摘要

In this paper, we compared the TL response of three types of thermoluminescence dosimeters, TLD600 (~6LiF:Mg,Ti), TLD700 (~7LiF:Mg,Ti) and MCP (LiF:Mg,Cu,P) after exposure to a n-γ mixed field in the fluence range of radiotherapeutic applications. Since a dosimeter pair is required to discriminate the two components of the mixed field, we analyzed the ability of each dosimeter pair to provide the fluence value in the mixed field. At this aim we performed a ~(60)Co-γ calibration and a neutron calibration for all three dosimeter types. Finally, a blind test was performed in order to analyze the accuracy of each dosimeter pair and we found that in this mixed field the fluence value obtained through the TLD600-MCP pair is as accurate as the value obtained through the most common TLD600-TLD700 pair.
机译:在本文中,我们比较了三种类型的热致发光剂量计的TL响应:TLD600(〜6LiF:Mg,Ti),TLD700(〜7LiF:Mg,Ti)和MCP(LiF:Mg,Cu,P)在放射治疗应用的注量范围内的n-γ混合场。由于需要剂量计对来区分混合场的两个分量,因此我们分析了每个剂量计对在混合场中提供注量值的能力。为此,我们对所有三种剂量计类型进行了〜(60)Co-γ校准和中子校准。最后,进行盲测试以分析每个剂量计对的准确性,我们发现在该混合场中,通过TLD600-MCP对获得的注量值与通过最常见的TLD600-TLD700对获得的注量值一样准确。 。