首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >Electronic background rejection in a new ultra-low background alpha-particle counter
【24h】

Electronic background rejection in a new ultra-low background alpha-particle counter

机译:新型超低背景Alpha粒子计数器中的电子背景抑制

获取原文
获取原文并翻译 | 示例

摘要

Low background alpha-particle (α) counting is required in the semiconductor industry, where α's produce single event errors. Industry road maps call for measuring α emissivities at 0.0005 α/cm~2/h, while current commercial counter backgrounds are 0.005 α/cm~2/h, a factor of 10 too high. This paper shows that by designing an ionization chamber so that ionization tracks collected from the sample have long risetimes while those collected from the anode have short risetimes, signal risetime analysis can distinguish a emanation location within the counter. Coupled with a guard electrode to reject tracks emitted from the counter sidewalls, the method can achieve backgrounds approaching 0.0001 α/cm~2/h, a factor of 20 lower.
机译:在半导体行业中,低背景α粒子(α)计数是必需的,其中α会产生单事件错误。工业路线图要求在0.0005α/ cm〜2 / h的条件下测量α发射率,而目前的商业柜台背景为0.005α/ cm〜2 / h,这是10倍的太大。本文表明,通过设计一个电离室,使从样品收集的电离轨迹的上升时间较长,而从阳极收集的电离轨迹的上升时间较短,信号上升时间分析可以区分计数器内的发射位置。结合保护电极以拒绝从相对侧壁发出的走线,该方法可以实现接近0.0001α/ cm〜2 / h的背景,降低了20倍。