HVE has tested the performance of its first compact 1 MV AMS system. The system is primarily designed for the measurement of ~(14)C, ~(10)Be and ~(26)Al, but additionally supports the measurement of heavy ions like iodine and plutonium. The tests confirmed the flexibility of a system that results from its 1 MV nominal terminal voltage, as it allows the use of various charge states for the analysis of e.g. ~(14)C, supports ~(10)Be analysis with good detection efficiency and permits the analysis of heavy ions like ~(129)I and Pu. For the analysis of ~(10)Be, the isobaric ~(10)B is removed by means of the nuclear charge dependent energy loss in an absorber foil in front of the electrostatic analyser. The results of the performance tests for ~(14)C-, ~(10)Be- and ~(26)Al-analyses are presented in this paper.
机译:HVE已测试了其首款紧凑型1 MV AMS系统的性能。该系统主要设计用于〜(14)C,〜(10)Be和〜(26)Al的测量,但是还支持测量重离子,例如碘和p。测试证实了系统的灵活性是由于其1 MV的标称端电压所致,因为它允许使用各种充电状态来分析例如〜(14)C,支持〜(10)Be分析,检测效率高,可分析〜(129)I和Pu等重离子。为了分析〜(10)Be,借助静电分析仪前面的吸收器箔中依赖于核电荷的能量损失,去除了等压〜(10)B。本文介绍了〜(14)C-,〜(10)Be-和〜(26)Al分析的性能测试结果。