首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >t of geometrical parameter's uncertainty of BIXS experimental setup for tritium analysis
【24h】

t of geometrical parameter's uncertainty of BIXS experimental setup for tritium analysis

机译:IX分析的BIXS实验装置的几何参数不确定性t

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, the effect of geometrical parameter's uncertainty of experimental setup in β-decay induced X-ray spectroscopy (BIXS) method has been studied. By varying deliberately the distance between the X-ray detector and the sample surface, the change of the tritium content and the tritium depth distribution obtained by the BIXS method has been observed. Both Monte Carlo simulations and experimental measurements were carried out for our BIXS experimental setup. It is found that the BIXS method will not give reliable tritium content and depth distribution of the sample if the sample-detector distance of the experimental setup is not measured accurately, i.e., the deviation of 1 mm of the sample-detector distance will cause ~10% error of total tritium content besides of the distortion of the shape of tritium depth distribution. This indicates that the geometrical parameters of the experimental setup in the BIXS method must be accurately determined, especially the internal configuration of the X-ray detector used in the BIXS method. The measurement of the internal configuration of an X-ray detector used in the BIXS method by an industrial computed tomography (CT) has been shown in this paper.
机译:本文研究了β-衰变诱导X射线光谱法(BIXS)中几何参数的不确定性对实验装置的影响。通过故意改变X射线检测器与样品表面之间的距离,可以观察到通过BIXS方法获得的content含量和and深度分布的变化。我们的BIXS实验装置均进行了蒙特卡洛模拟和实验测量。如果未正确测量实验装置的样品检测器距离,则发现BIXS方法将无法提供可靠的tri含量和样品深度分布,即,样品检测器距离1 mm的偏差会导致〜除了tri深度分布形状的变形之外,总tri含量的误差为10%。这表明必须精确确定BIXS方法中实验装置的几何参数,尤其是BIXS方法中使用的X射线探测器的内部配置。本文显示了通过工业计算机断层扫描(CT)对BIXS方法中使用的X射线检测器的内部配置进行测量。

著录项

  • 来源
  • 作者单位

    Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China;

    Key Laboratory of Radiation Physics and Technology of Ministry of Education, Institute of Nuclear Science and Technology, Sichuan University, Chengdu 610064, China;

    Key Laboratory of Radiation Physics and Technology of Ministry of Education, Institute of Nuclear Science and Technology, Sichuan University, Chengdu 610064, China,Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China;

    Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China;

    National X-ray Digital Imaging Instrument Center, Mianyang 621900, China;

    Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    tritium analysis; monte carlo simulation; tikhonov regularization;