Improved stability of multi-reflection time-of-flight mass spectrometers through passive and active voltage stabilization

机译:通过无源和有源电压稳定度提高了多反射飞行时间质谱仪的稳定性

获取原文
获取原文并翻译 | 示例

摘要

Multi-reflection time-of-flight (MR-ToF) spectrometers are devices in which ions are reflected between two electrostatic mirrors to prolong their flight path. The performance of the apparatus relies on the stability of the voltages that are used to define the mirror potentials. Especially the voltage of the mirror electrodes at the point where the ions reverse the direction of their axial motion needs to be stable in order to minimize fluctuations of their total flight time. Here we present a method to increase the short- and long-term stability of suitable voltage supplies for enhanced performance of the spectrometer.
机译:多反射飞行时间(MR-ToF)光谱仪是这样的设备,其中离子在两个静电镜之间反射,以延长其飞行路径。设备的性能取决于用于定义镜面电势的电压的稳定性。特别地,在离子反转其轴向运动方向的点处的镜电极的电压需要稳定,以便最小化其总飞行时间的波动。在这里,我们提出了一种方法,用于增加合适电压源的短期和长期稳定性,以提高光谱仪的性能。