首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Highly sensitive measurement of α-rays in NaI(Tl) scintillator by pulse shape discrimination with charge-sensitive analog-to-digital converter
【24h】

Highly sensitive measurement of α-rays in NaI(Tl) scintillator by pulse shape discrimination with charge-sensitive analog-to-digital converter

机译:电荷敏感型模数转换器通过脉冲形状识别对NaI(Tl)闪烁体中的α射线进行高灵敏度测量

获取原文
获取原文并翻译 | 示例

摘要

In order to investigate the contamination due to the radioactive isotopes (RIs) in NaI(Tl) scintillator for dark matter search, a simple method of pulse shape discrimination (PSD) to extract α-rays energy spectrum was developed. Two linear signals of different timings were input into a single module charge-sensitive analog-to-digital converter. The α-and the γ-rays were discriminated by comparing the total charge integration of a pulse with the partial one. As a result, the α-rays energy spectrum with NaI(Tl) was successfully and clearly extracted. The present work showed that the PSD method was performed by the minimum electronic circuits. The sensitivity was high enough to measure RIs of a few μBq/kg in high pure NaI crystals for dark matter search.
机译:为了研究NaI(Tl)闪烁体中放射性同位素(RIs)进行暗物质搜索所造成的污染,开发了一种简单的脉冲形状识别(PSD)提取α射线能谱的方法。将具有不同时序的两个线性信号输入到单个模块的电荷敏感型模数转换器中。通过比较脉冲与部分脉冲的总电荷积分来区分α射线和γ射线。结果,成功且清晰地提取了具有NaI(Tl)的α射线能谱。目前的工作表明,PSD方法是由最少的电子电路执行的。灵敏度足够高,可以测量高纯度NaI晶体中用于暗物质搜索的RIs为几μBq/ kg。